期刊文献+

流动脉动对流量测量仪表影响的几个实例 被引量:11

The Examples as the Effect of Flow Pulsation on Flow Measurement Instrumentation
下载PDF
导出
摘要 流量测量中经常遇到脉动流,它对容积式流量计之外的绝大多数种类的流量计产生影响,使之误差增大,将脉动源消除或在流量计与脉动源之间装上阻尼器,能有效克服这种影响。列举了3个实例,流体有液体也有蒸汽;仪表有差压流量计也有旋涡流量计。 During the flow rate measurement,the flow pulsation is often found,which could affect most types of the measurement instruments beside the volumetric meter and cause the error. There are two ways to deal with this problem. The first one is to cut off the resource of the flow pulsation,and the second one is to install a damper between flowmeter and the resource of flow pulsation. Both of ways mentioned above can effectively get over this problem. Three examples are listed. The fluid could be liquid or steam. There are two types of flowmeter,differential pressure type and vortex type.
作者 宋文伟 纪纲
出处 《石油化工自动化》 CAS 2003年第5期75-78,84,共5页 Automation in Petro-chemical Industry
关键词 流量测量仪表 流动脉动 孔板流量计 旋涡流量计 差压流量计 flow pulsation flowmeter damper orifice-plate flowmeter vortex flowmeter
  • 相关文献

参考文献6

  • 1米勒RW.流量测量工程手册[M].北京:机械工业出版社,1990.637-644.
  • 2米勒R W 孙延祚译.流量测量工程手册[M].北京:机械工业出版社,1990..
  • 3.GB/T 2624-93.流量测量装置用孔板、喷嘴和文丘里管测量充满圆管的流体流量[S].,..
  • 4ISO/TR 3313,1998, Measurement of fluid flow in closed conduits--Guidelines on the effects of flow pulsations on flow-measurement instruments[S].
  • 5ISO 5167-1991, Measurement of Fluid Flow by Means of Orifice Plates, Nozzles or Venturi Tubes Inserted in Circutar Cross-Section Conduits Running Full[S].
  • 6GB/T2624—93.流量测量装置用孔板、喷嘴和文丘里管测量充满圆管的流体流量[S].[S].,..