摘要 采用对低能X射线具有高探测效率和高分辨率的HPGe探测器,用和峰法测量了139Ce溶液的比活度,测量的结果为762.9(1±0.5%)Bq/mg,,国内比对推荐值为765.9(1±0.6%)Bq/mg,二者在不确定度范围内符合得很好。 Based on HPGe detector with high efficiency and resolution for low energy Xray, we have measured the specific activity of 139Ce by sumpeak method. The result obtained is 762.9(1±0.5%)Bq/mg, which is well agreed with recommended result 765.9(1±0.6%)Bq/mg of native compared measurement in uncertainty.
出处 《核电子学与探测技术》 CAS CSCD 北大核心 2003年第5期448-450,共3页 Nuclear Electronics & Detection Technology
关键词 HPGEΓ谱仪 ^139Ce 活度 峰效率 核素 HPGE探测器 ^(139)Ce HPGe detector sum-peak method activity
  • 相关文献

参考文献3

  • 1Wang Zhonglu. The cetermmation oi 125I activity using sum-peak method with a well-type HPGe-derector-based spectrometer [J]. Nucl Instrum Methods,2001, A459:471.
  • 2Korun D. Coincedence summing in gamma and X- ray spectrometry [J]. Nucl Instrum Methods, 1993,A325:478.
  • 3Richard B Firestone. Table of isotopes CD-ROM [M]. Eighth Edition, 1996.