摘要
目的:探讨适合国内患者的人工晶体屈光度计算的A常数值。方法:统计分析了355只后房型人工晶体眼的生理参数,并用二元线性回归对355只后房型人工晶体眼进行A常数计算。结果:康明、Storz和Pharmacia人工晶体的A常数分别为116.9、117.6和117.8,回归标准分别为2.44、1.83和2.98。结论:利用国内患者的临床资料进行二元线性回归而得的A常数值可指导临床医师更准确地计算人工晶体屈光度值。
Objective: To probe into the constant A which is suitable for the calculation and correction of it
among internal patients with pseudophakia. Methods: Statistical analysis of physiological
parameters obtained from 355 patients with pseudophakia was made and the constant A among
355 patients with pseudophakia was calculated with dual linear regression. Results: The
constants A of KM,Storz and Pharmacia were 116.9,117.6 and 117.8 respectively while standard
deviation was 2.44,1.83 and 2.98 respectively. Conclusion: The constant A obtained by
exploiting clinic data gathered from internal patients can be helpful to those clinic doctors in
calculating the diopter of IOL more accurately.