期刊文献+

改善密封源仪表安全状况的建议 被引量:1

Suggestions to Improve Safety of Sealed Radioactive Source-inbuilt Devices
下载PDF
导出
摘要 探讨了改善密封源仪表应用安全状况的措施。通过理顺审管体制 ,提高监管质量 ,强化运营管理责任 ,倡导安全文化等措施 。 An approach is made to improve the safety of devices in which sealed radioactive sources inbuilt. With a view to attainting the goal of safe application of the devices/meters equipped with sealed radioactive sources, a wide range of suggestions are put forth, including streamlining regulatory system, improving oversight and control quality, reinforcing the responsibilities for operation and management and advocating safety culture.
出处 《辐射防护通讯》 2003年第6期24-26,共3页 Radiation Protection Bulletin
关键词 密封源仪表 安全管理 安全文化 中国 放射源事故 Device containing sealed radioactive source Safety Suggestion of improvement
  • 相关文献

参考文献3

二级参考文献21

  • 1林殿科.伽玛刀治疗室混凝土屏蔽厚度的设计[J].辐射防护通讯,1996,16(1):22-26. 被引量:1
  • 2鲍首琛.放射治疗与核安全文化[J].辐射防护通讯,1996,16(6):14-27. 被引量:4
  • 3卫生部卫生法制与监督司.1997年卫生监督工作情况通报[M].,1998,4..
  • 4卫生部卫生法制与监督司.1998年卫生监督工作情况通报[M].,1999,4..
  • 5孙文钧 俞荣生.10次倒装钴源的现场监督体会[J].中国辐射卫生,1997,6(4):213-213.
  • 6李天来 张居政 等.应重视对刻度仪器用的弱放射源的防护管理[J].中国辐射卫生,1998,7(2):181-181.
  • 7孙健 汪春亮.小型密封放射源安全贮存的几点做法[J].中国辐射卫生,1998,7(4):245-245.
  • 8谢华 范辉堂.一起丢失放射源事故的处理与分析[J].中国辐射卫生,1998,7(2):112-112.
  • 9徐开宇 尤建国 等.常州市丢失放射源事故应急处理学习的体会[J].中国辐射卫生,1998,7(1):20-20.
  • 10李显明 卓维海 等.一起源辐照装置降源反卷事件及其处理[J].中国辐射卫生,1997,6(2):115-115.

共引文献15

同被引文献16

引证文献1

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部