摘要 目的:探讨碱性清洗剂联合超声清洗机清洗妇科复用器械的效果。方法:选择吸引头、窥阴器共300个,随机分成A、B、C 3个组,A组用医用软毛刷刷洗后用高压水枪冲洗,然后放入全自动清洗机清洗;B组用p H值中性的强效多酶清洗剂浸泡10 min后再用A组方法清洗;C组放入已加碱性清洗剂的超声机震动10 min再用A组方法清洗,比较清洗效果。结果:清洗后,A组合格率58%,B组合格率90%,C组合格率100%。结论:采用超声机震动+碱性清洗剂+毛刷+高压水枪冲洗,其清洗消毒效果最佳。 Objective:To explore the effect of alkaline cleaning agent combined with ultrasonic cleaning machine in cleaning equipment in gynecology.Methods:300 suction head and vaginal speculum were selected.They were randomly divided into A group,B group,C group.A group was given the medical soft brush washing and rinsing with water cannons,and then they were putted into the automatic cleaning machine.B group was soaked for 10 minutes with pH neutral potent multi enzyme cleaning agent,and then washed with A method.C group was shaked for 10 minutes in the ultrasonic vibration machine with alkaline cleaning agent,and then washed with A method.We compared the cleaning effect.Results:After cleaning,the qualified rate of A group was 58%;the qualified rate of B group was 90%;the qualified rate of C group was 100%.Conclusion:Cleaning effect was the best of ultrasound machine vibration plus alkaline cleaning agent plus high-pressure water washing brush.
作者 冉茂勤
出处 《中国社区医师》 2015年第19期129-129,132,共2页 Chinese Community Doctors
关键词 碱性清洗剂 超声震动 妇科吸引头 清洗效果 Alkaline cleaning agent Ultrasonic vibration Gynecological suction head The cleaning effect
  • 相关文献

参考文献5

二级参考文献38

  • 1余如平.腹腔镜手术器械清洗消毒中的问题与对策[J].中华医院感染学杂志,2005,15(4):420-420. 被引量:39
  • 2宋烽,董薪,许多朵.手术器械清洗方法的研究[J].中华医院感染学杂志,2006,16(4):410-411. 被引量:99
  • 3李大忻.医疗器械清洗前后效果比较[J].中华医院感染学杂志,2006,16(5):545-545. 被引量:11
  • 4贾莉,靳明萍,任绪华,马为梅.腹腔镜精密器械清洗方法的研究[J].护理学杂志(外科版),2006,21(9):53-54. 被引量:10
  • 5中国卫生部.医院消毒供应中心第3部分:清洗消毒及灭菌效果监测标准.2009.4.
  • 6Alfa MJ,De Gagne P,Olson N,et al. Comparison of ion plas- ma,vaporized hydrogen peroxide, and 100~ ethylene oxide sterilizers to the 12 /88 ethylene oxide gas sterilizer[J]. In- fect Cont Hosp Epidemiology, 1996,17(1) :92-93.
  • 7Fraser VJ, Zuckerman G, Clouse RE, et al. A prospective randomized trial comparing manual and automated endoscope disinfection methods[J]. Infect Cont Hosp Epidemiology, 2003,14 (4) L383-385.
  • 8Vyas A, Pickering C, Oldham L,et al. Survey of symptoms, respiratory function,and immunology and their relation to gl- utaraldehyde and other occupational exposures among endo- scope nursing Employee[J]. Occup Environ Med, 2000, 57 (7) :752-753.
  • 9Oulahal LagsirN. Ultrasonic methodology coupled to ATP bi- oluminescence for the noninvasive detection of fouling in food processing equipment-validation and application to a dairy fac- tory[J]. Journal of applied microbiology, 2000, 89: 433.
  • 10华九月,王桂娣,郑润香,陈洁.手术器械清洗效果的检测方法[J].中华护理杂志,2008,43(7):671-672. 被引量:57

共引文献52

同被引文献25

引证文献3

二级引证文献4

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部