期刊文献+

内置式膜式氧合器模型氧合效果的评价 被引量:6

Evaluation of Oxygen Transfer Rate of Intravenous Membrane Oxygenator Modules
原文传递
导出
摘要 设计了三种不同编排方式的静脉内中空纤维膜式氧合器(IVOX)模型,用Na2SO3的水溶液进行了氧合器的水气传质实验,以Na2SO3的氧化反应来模拟血红蛋白的氧合,并且建立了符合此体系的半经验数学模型,求算相应的血氧交换速率.结果表明,纤维的编排方式对氧合器的氧合效果有显著的影响,网线式的编排具有最好的氧合效果. Oxygen transfer rate of intravenous membrane oxygenator (IVOX) modules made of polypropylene hollow fibers were evaluated in vitro experiments using Na_(2)SO_(3) solution instead of blood. A semi-empirical mathematical model was established to predict oxygen- blood transfer rates in IVOX. The results showed that the modules geometry and packing of hollow fibers affected the transfer rates remarkably. The IVOX module with reticulated woven fibers had the highest blood-oxygen transfer rate.
出处 《复旦学报(自然科学版)》 CAS CSCD 北大核心 2003年第6期979-983,共5页 Journal of Fudan University:Natural Science
基金 973"国家重点研究专项经费资助项目(G1999064800)
关键词 内置式膜式氧合器(IVOX) 微孔聚丙烯膜 中空纤维 血氧交换速率 人工肺 intravenous membrane oxygenator (IVOX) microporous polypropylene membrane hollow fibers blood-oxygen transfer rate
  • 相关文献

参考文献2

二级参考文献4

共引文献2

同被引文献117

  • 1江朝光,林刚,李功宋,周津,朱朗标,杨溥臣.血管内氧合器的研制[J].中华胸心血管外科杂志,1995,11(4):234-235. 被引量:3
  • 2江朝光,林刚,李功宋,周津,杨溥臣,马莉.血管内氧合器的研制及实验研究[J].中华医学杂志,1995,75(9):561-562. 被引量:2
  • 3谢钢,蒋崇慧,李斌飞,宁晔,张志刚,姜海明,郑伟华,尹刚,赵双彪.体外膜肺氧合在严重肺挫伤中的应用[J].中国综合临床,2005,21(9):817-819. 被引量:24
  • 4许少波,王建华.中空纤维膜式氧合器的研究进展[J].医疗保健器具,2006,13(10):4-7. 被引量:6
  • 5Hattler B G, Federspiel W J. Gas exchange in the venous system: support for the failing lung. In: Vaslef SN,Anderson RW, eds. The artifieial lung [M]. Austin TX: Landes Bioscienee, 2002.
  • 6Hill J D, O'Brien T G, Murray J J,et al. Prolonged extracorporeal oxygenation for acute post - traumatic respiratory failure (shock - lung syndrome) .Use of the Bramson membrane lung [J]. N Engl J Med, 1972,286(12):629-634.
  • 7Arpad P,Georg L,Heinz S,et al. Bilateral lung transplantation with intra and postoperatively prolonged ECMO support in patients with pulmonary hypertension [J]. European Journal of Cardio-thoracic Surgery, 2002,21 (5) : 858.
  • 8Mortensen J D. An intravenacaval blood gas exchange (IVCBGE) device, a preliminary report [J]. ASAIO Trans, 1987,33 ( 3 ) : 570- 573.
  • 9Kallis P,Al Saady N M,Bennett E D,et al. Early results of intravascular oxygenation [J]. Eur J Cardiothorac Surg, 1993,7 (4): 206-210.
  • 10Kanamori T,Niwa M,Kawakami H,et al. Estimate of gas transfer rates of an intravascular membrane oxygenator [J]. ASAIO J, 2000,46 (5) :612-619.

引证文献6

二级引证文献7

相关作者