期刊文献+

CPAP呼吸机固定压力与自调压力在同组OSAS患者治疗中的初步比较

Comparison between CPAP Ventilator and Auto-CPAP in the Treat ment of Patients with OSAS
下载PDF
导出
摘要 目的 比较两种CPAP呼吸机在治疗过程中压力变化与否对OSAS患者治疗的优缺点。方法  18例被诊断为OSAS并接受CPAP治疗的患者首先接受自调压力呼吸机治疗 ,并据该结果 ,按最佳氧饱和度及AHI时的压力水平 ,接受固定压力呼吸机的治疗。同时 ,对比评判其AHI ,SaO2 ,心率等的变化。结果 自调压力呼吸机在OASA患者治疗时 ,其变化的压力对AHI、SaO2 、心率均构成影响 ,从病人生理指标来看 ,固定压力呼吸机可能对病人治疗更为有利 。 Objective When we use the CPAP ventilator to treat th e patients with Obstructive Sleep Apnea syndrome (OSAS),the optimal pressure req uired is usually established in the sleep laboratory by doctor.Recently,there ar e self-adjusting CPAP machines(auto-CPAP) depend on computer controling.This p resent study was designed to evaluate these two kind machines in treatment of va rious pressure.Methods Following this study,18 patients with OSAS w ere first allocated to reseive arto-CPAP treatment under control.And then,accor ding to this treatment result,we select the conventional pressure by the best AH I and SaO 2 for these patients again.We did the observation of their AHI,SaO 2,heart rate and so on.Results When we use auto-CPAP,their AHI, SaO 2,heart rate were changed with various pressure.We suggest that convention al CPAP may be far better than auto-CPAP,and auto-CPAP ventilator is the tool for doctor to titrate the conventional pressure.
出处 《中国血液流变学杂志》 CAS 2003年第4期414-415,共2页 Chinese Journal of Hemorheology
关键词 CPAP呼吸机 固定压力 自调压力在 OSAS 治疗 CPAP ventilator conventional pressure auto-CPAP OSAS