期刊文献+

冠状动脉追踪冻结技术和冠脉DSA成像的比较研究

下载PDF
导出
摘要 目的研究冠状动脉追踪冻结技术(SSF)采用前瞻性心电门控扫描与同期DSA成像的比较。方法共纳入80例行SSF和DSA成像的冠心病患者,平均心率(75.6±8.2)次/min,均不服用降低心率药物,支架置入29例(36.25%)。SSF采用美国GE 64排128层冠脉Optima CT680扫描机和GE AW4.6 Advantage workstation后处理工作站,前瞻性心电门控扫描,比较SSF与DSA成像评估冠脉粥样斑块和管腔狭窄程度的价值。结果 SSF成像各组不同心率显示斑块数目、管腔狭窄程度以及管腔狭窄≥75%数目比较均无差异(P>0.05);DSA与SSF成像在≤65次/min和65~85次/min组管腔狭窄程度以及管腔狭窄≥75%数目比较无差异(P>0.05),但是>85次/min组DSA成像较SSF显示管腔狭窄程度以及管腔狭窄≥75%数目明显增加,差异有统计学意义(P<0.05)。结论在心率<85次/min SSF采用前瞻性心电门控扫描与同期DSA成像在显示粥样斑块数目和管腔狭窄程度上价值相当。
出处 《中国医疗设备》 2017年第S2期134-134,共1页 China Medical Devices
基金 湖北省自然科学基金立项课题 课题编号:2016CFB697
  • 相关文献

参考文献4

二级参考文献29

  • 1Maffei E,Martini C,De Crescenzo S. Low dose CT of the heart:a quantum leap into a new era of cardiovascular imaging[J].La Radiologia Medica,2010.1179-1207.
  • 2Hsieh J,Londt J,Vass M. Step-and-shoot data acquisition and reconstruction for cardiac x-ray computed tomography[J].Medical Physics,2006.4236-4248.
  • 3Achenbach S, Ropers U, Kuettner A, et al. Randomized Comparison of 64-slice Single-and-dual-Source Computed tomography coronary angiography for the detection of coronary artery disease 1. Jacc Cardiovascular imaging, 2008 ,3 : 177 -186.
  • 4Mir-Akbari H,Ripsweden J,Jensen J,et al. Limitations of 64-detector-row computed tomography coronary angiography: calcium and motion but not short experience J Acta Radiol ,2009,50: 174-180.
  • 5Leipsic J, Labounty TM, Hague CJ, et al. Effect of a novel vendor specific motion-correction algorithm on image quality and diagnostic accuracy in persons undergoing coronary CT angiography without rate control medications. J Cardiovascular Compute Tomography,2012, 6: 164-171.
  • 6Ghadri JR, Kuest SM, Goetti R, et al. Image quality and radiation dose comparison of prospectively triggered low-dose CCTA: 128- slice dual-source high-pitch spiral versus 64-slice single-source sequential acquisition. Int J Cardiovascular Imaging, 2011 ,28: 1217- 1225.
  • 7Paul JF, Amato A, Rohnean A. Low-dose coronary-CT angiography using step and shoot at any heart rate: comparison of image quality at systole for high heart rate and diastole for low heart rate with a 128- slice dual-source machine[J]. Int J Cardiovasc Imaging, 2013 ,29 : 651-657.
  • 8Apfaltrer P, Schoendube H, Schoepf UJ, et al. Enhanced temporal resolution at cardiac CT with a novel CT image reconstruction algorithm: initial patient experience. Eur J Radiol ,2013 ,82 :270-274.
  • 9Machida H, Lin XZ, Fukui R, et al. Influence of the motion correction algorithm on the quality and interpretability of images of singlesource 64-detector coronary CT angiography among patients grouped by heart rate. Jpn J Radiol,2015 ,33 :84-93.
  • 10Fuchs TA, Stehli J, Dougoud S, et al, Impact of a new motion-correction algorithm on image quality of low-dose coronary CT angiography in patients with insufficient heart rate control. J Acad Radiol, 20 14 , 21 :312-317.

共引文献34