摘要
建立了测定布洛芬含量的高效液相色谱法。选用Agilent Eclipse XDB-C8柱(4.6mm×250mm,5μm),LabAlliance C8保护柱(4.6mm×10mm),流动相为甲醇:水=75:25,流速1.0mL/min,检测波长220nm,进量样20μL,柱温:室温。在10.0~400μg/mL浓度范围内,线性方程为A=52.50C+50.79,相关系数0.999 9,回收率96.0%~100.7%,RSD0.65%~1.32%。该方法简便、准确、快速,可用于布洛芬的质量控制。
Ibuprofen sustained release capsules was analysed by HPLC method, with Agilent Eclipse XDB-C8 col-umn(4.6 mm× 250 mm,5 μm) . The mobile phase was mixture of methanol and water(75 :25) .UV detector was applied with 220 nm. The average recovery was 96.0%-100.7% ,with RSD of 0.65%-1.32% (n = 15) .The method is simple, rapid and the results is accurate and reliable.
出处
《浙江科技学院学报》
CAS
2003年第B11期90-91,共2页
Journal of Zhejiang University of Science and Technology