摘要 为评估膜部室间隔缺损(室缺)介入治疗的效果,回顾性总结21例膜部室缺封堵术的临床应用。超声心动图显示膜部室缺直径4~8mm。选用Judkin右冠脉导管和泥蚯导丝,通过室缺建立轨道,采用Amplatzer膜部室缺封堵器封堵。结果表明,20例治疗成功,1例嵴内型室缺术后封堵器脱落。20例随访1~8个月,无异常表现。结论:在严格选择适应证和熟练操作技巧的条件下,膜部室缺封堵术是一项操作安全、疗效可靠的方法,值得推广应用。
出处 《沈阳部队医药》 2004年第1期5-7,共3页
  • 相关文献

参考文献4

  • 1Arora R,,Trehan V,Kumar A, et al.Transcatheter closure of congenital ventricular septal defects: experience with various devices[].Interv Cardiol.2003
  • 2Hijazi ZM,,Hakim F,Haweleh AA, et al.Catheter closure of perimembranous ventricular septal defects using the new Amplatzer membranous VSD occluder: initial clinical experience[].Catheterization and Cardiovascular Interventions.2002
  • 3Bass JL,Kalra GS,Arora R,et al.Initial human experience with the amplatzer perimembranous ventricular septal occluder device[].Catheterization and Cardiovascular Interventions.2003
  • 4Gu XP,Han YM,Titw S,et al.Transcatheter closure of membranous ventricular septal defects with a new nitinol prosthesis in a natural swine model[].Catheter Cardiovasc Interv.2000