期刊文献+

测定锰(Ⅱ)离子的KIO_4-K_2CO_3化学发光体系 被引量:6

Flow injection chemiluminescence determination of manganese(Ⅱ) by using KIO_4-K_2CO_3 system
下载PDF
导出
摘要 碱性介质中 ,KIO4与二价锰离子反应 ,在没有任何化学发光试剂存在条件下产生微弱的化学发光。当向体系中加入碳酸盐后 ,反应产生的化学发光显著增强 ,且该发光强度与Mn2 + 浓度在 1× 1 0 -8~ 5× 1 0 -6mol L范围内成线性关系 ,测定检出限为 5× 1 0 -9mol L。基于这一现象 ,同时结合流动注射分析技术 ,建立起了一种测定环境水中微量Mn2 + In basic solution, the mixture of periodate and carbonate showed a weak chemiluminescence. When a trace amount of manganese(Ⅱ) was injected into the mixture solution, a strong chemiluminescence can be recorded. Based on this, a flow injection chemiluminescence system, Mn(Ⅱ)-KIO_4-K_2CO_3 without luminescent reagents, was assembled to determine manganese(Ⅱ) in water. The CL emission intensity had a linear relationship with the concentration of manganese(Ⅱ) in the range of 1.0×10^(-8)~5×10^(-6) mol/L. The relative standard deviation for 1.0×10^(-6) mol/L manganese(Ⅱ) was 2.7% (n=9). The detection limit was 5×10^(-9) mol/L (S/N=3). This method was used to determine manganese in well water with the recoveries of 98%~102%.
出处 《分析试验室》 CAS CSCD 北大核心 2004年第1期49-53,共5页 Chinese Journal of Analysis Laboratory
基金 国家杰出青年科学基金 (2 0 12 5514 ) 中国科学院"百人计划"资助项目
关键词 锰(Ⅱ)离子 含量测定 KIO4-K2CO3化学发光体系 痕量分析 流动注射化学发光分析 Chemiluminescence Flow injection analysis Periodate Manganese