期刊文献+

等速肌张力测试系统的开发 被引量:2

Development of Isokinetic Muscle Tone Testing System
下载PDF
导出
摘要 针对帕金森病患者的肌张力障碍,该文设计了一种新型等速肌张力测试系统。该系统包括机械装置、数据采集单元和上位机数据处理分析软件。其中机械装置包括手臂支撑装置、安全限位装置、高度调节装置等;数据采集单元由力矩传感器模块、电机控制模块、数据通信模块等组成,用于控制手臂进行等速运动并将传感器的模拟信号转换成数字信号实时传输到上位机;上位机数据处理分析软件采用USB接口传输数据,基于VC++程序开发,完成数据处理、显示、存储及报表生成等功能。实验结果表明:设计的等速肌张力测试系统安全可靠,不仅能够较好地评估帕金森患者的肌张力异常程度,还能应用于辅助肌张力患者的康复训练,具有较强的可扩展性。 A new type of isokinetic muscle tone testing system was designed for the Parkinson’s patients with dysmyotonia. The system consists of a mechanical device, a data acquisition unit, and a data processing and analyzing software. The mechanical device comprises an arm supporting device, a safety limiting device and a height adjusting device. Data acquisition unit, composed of a torque sensor module, a motor control module, a data communication module, and so on, is used to control the arm at constant speed and convert the sensor’s analog signal into digital signal, then the signal is transmitted to the computer in real-time by USB interface. The data processing and analyzing software, developed by VC++, has finished the data processing, display, storage, generated the reports, et al. The experimental results show that this system is reliable and safe, not only can evaluate the degree of abnormal muscle tone in Parkinson’s patients, but also can be used in the rehabilitation training of patients with dysmyotonia, which has strong expansibility.
出处 《中国医疗器械杂志》 2016年第3期157-160,共4页 Chinese Journal of Medical Instrumentation
基金 国家科技支撑计划(2013BAH14F01) 安徽省科技攻关计划课题(1301042215 1501021042)
关键词 帕金森病 肌张力障碍 等速测量 USB通信 Parkinson dysmyotonia isokinetic testing USB communication
  • 相关文献

参考文献13

  • 1胡德福.应变式扭矩传感器的设计技术[J].船舶工程,2011,33(4):96-99. 被引量:24
  • 2黄志平,尹彦,刘敏,黄中校,杨小英,苏玉林.等速肌力测试与训练技术的研究进展[J].体育科技,2011,32(4):52-58. 被引量:53
  • 3徐秀林,蒋奇永,严荣国,朱政康,郭旭东.基于蓝牙通信的等长肌力测试仪的设计[J].中国生物医学工程学报,2011,30(6):897-903. 被引量:6
  • 4林己,廖一鹏.基于STM32的伺服电机控制方法研究与设计[J].工业控制计算机,2013,26(3):4-5. 被引量:8
  • 5Zawadzki J,Bober T,Siemienski A.Validity analysis of the Biodex System 3 dynamometer under static and isokinetic conditions. Acta Bioeng Biomech . 2010
  • 6Wright David,Nakamura Kazuhiro,Maeda Tetsuya,Kutsuzawa Keiichi,Miyawaki Kazuhito,Nagata Ken.Research and development of a portable device to quantify muscle tone in patients with Parkinsons disease. Conference proceedings : ... Annual International Conference of the IEEE Engineering in Medicine and Biology Society. IEEE Engineering in Medicine and Biology Society. Conference . 2009
  • 7Blackburn Marjan,van Vliet Paulette,Mockett Simon P.Reliability of measurements obtained with the modified Ashworth scale in the lower extremities of people with stroke. Physical Therapy . 2002
  • 8Higher Muscle Passive Stiffness in Parkinson’s Disease Patients Than in Controls Measured by Myotonometry(J)Archives of Physical Medicine and Rehabilitation . 2010 (5)
  • 9WalterMaetzler,Jeffrey M.Hausdorff.??Motor signs in the prodromal phase of Parkinson’s disease(J)Mov. Disord. . 2012 (5)
  • 10王学伟,张未未.USB—GPIB控制器VISA函数库的开发及在Visual C++中的应用[J].电测与仪表,2006,43(11):50-53. 被引量:14