一体化同步扫描TOF-PET/MR分子影像设备的进展 被引量:4

Technical Progress of Integrated Simultaneous TOF-PET/MR Molecular Imaging Equipment
下载PDF
导出
摘要 该文介绍了一体化、同步扫描飞行时间正电子发射断层/磁共振分子影像设备的原理、结构和技术进展。简要介绍了一体化PET/MR设备的设计和组成的分类。探讨了一体化PET/MR中TOF技术发挥的作用和临床价值,实现TOF技术对PET探测器中光电转化组件的要求和完成PET/MR同步扫描的必要条件。比较了固态阵列光电转换器和雪崩光电二极管的性能差异。简单介绍目前一体化TOF-PET/MR设备采用MR成像参数对PET图像衰减校正面临的问题和技术挑战。 This paper introduces component, principle and technical progress of integrated simultaneous TOF-PET/MR molecular imaging equipment. It briefly describes the integrated PET/MRI device design and composition classification. It discusses special contributions to clinical application and research with TOF-PET/MR, requirement of photoelectric component on PET detector for attaining TOF reconstruction and the prerequisite that performed simultaneous scan on PET/MR. It compares the performance differences between APD and SSPM. The problems and the challenges of PET attenuation correction with MR image parameters are introduced briefly.
出处 《中国医疗器械杂志》 2016年第4期267-270,274,共5页 Chinese Journal of Medical Instrumentation
基金 山西省卫生厅科研课题(2015054)
关键词 飞行时间 一体化TOF-PET/MR PET衰减校正技术 磁共振 TOF TOF-PET/MR PET attenuation correction MR
  • 相关文献

参考文献10

  • 1Hofmann, Matthias,Steinke, Florian,Scheel, Verena,Charpiat, Guillaume,Farquhar, Jason,Aschoff, Philip,Brady, Michael,Sch?lkopf, Bernhard,Pichler, Bernd J.??MRI-Based Attenuation Correction for PET/MRI: A Novel Approach Combining Pattern Recognition and Atlas Registration(J)The Journal of Nuclear Medicine . 2008 (11)
  • 2Zaidi Habib.Is MR-guided attenuation correction a viable option for dual-modality PET/MR imaging?. Radiology . 2007
  • 3Kolb, Armin,Lorenz, Eckart,Judenhofer, Martin S,Renker, Dieter,Lankes, Konrad,Pichler, Bernd J.Evaluation of Geiger-mode APDs for PET block detector designs. Physics in Medicine and Biology . 2010
  • 4赵周社,辛军,郭启勇,王爽,陆喆.MRI弥散加权成像和PET/CT扫描在肿瘤检查应用的进展[J].中国临床医学影像杂志,2010,21(6):413-418. 被引量:18
  • 5赵周社,辛军,郭启勇,王爽,吴君.多种正电子示踪剂PET联合显像在肿瘤基础研究、临床诊断和疗效监测中的应用进展[J].中国临床医学影像杂志,2009,20(10):764-769. 被引量:5
  • 6Keereman, Vincent,Fierens, Yves,Broux, Tom,De Deene, Yves,Lonneux, Max,Vandenberghe, Stefaan.??MRI-Based Attenuation Correction for PET/MRI Using Ultrashort Echo Time Sequences(J)The Journal of Nuclear Medicine . 2010 (5)
  • 7Martinez-M?ller, Axel,Souvatzoglou, Michael,Delso, Gaspar,Bundschuh, Ralph A,Chefd’hotel, Christophe,Ziegler, Sibylle I,Navab, Nassir,Schwaiger, Markus,Nekolla, Stephan G.??Tissue Classification as a Potential Approach for Attenuation Correction in Whole-Body PET/MRI: Evaluation with PET/CT Data(J)The Journal of Nuclear Medicine . 2009 (4)
  • 8Matthias Hofmann,Bernd Pichler,Bernhard Sch?lkopf,Thomas Beyer.??Towards quantitative PET/MRI: a review of MR-based attenuation correction techniques(J)European Journal of Nuclear Medicine and Molecula . 2009 (1)
  • 9Gaspar Delso,Sibylle Ziegler.??PET/MRI system design(J)European Journal of Nuclear Medicine and Molecular Imaging . 2009 (1)
  • 10Schlemmer Heinz-Peter W,Pichler Bernd J,Schmand Matthias,Burbar Ziad,Michel Christian,Ladebeck Ralf,Jattke Kirstin,Townsend David,Nahmias Claude,Jacob Pradeep K,Heiss Wolf-Dieter,Claussen Claus D.Simultaneous MR/PET imaging of the human brain: feasibility study. Radiology . 2008

二级参考文献61

  • 1丁建平,王霄英,周良平,王继琛,肖江喜,蒋学祥.正常前列腺和精囊的磁共振弥散加权成像初步研究[J].中国医学影像技术,2004,20(8):1172-1174. 被引量:21
  • 2胡春洪,杨玲,陈剑华,丁乙.大脑星形细胞瘤ADC值与1HMRS主要化合物浓度的相关性[J].中国CT和MRI杂志,2005,3(1):5-7. 被引量:7
  • 3李德军,包尚联,马林.不同b值和扩散张量成像导出量的定量关系研究[J].中华放射学杂志,2004,38(12):1238-1242. 被引量:26
  • 4吴冰,张挽时,王东,孟利民,宋云龙,熊明辉,时惠平.MR DWI监测兔肝VX2肿瘤的实验研究[J].放射学实践,2006,21(3):240-243. 被引量:3
  • 5Jim B.FDG PET Reimbursement[J].J Nucl Med Technol,2004,32(1):33-38.
  • 6Avril N,Rose CA,Schelling M,et al.Breast imaging with positron emission tomography and fluorine-18 fluorodeoxyglucose:use and limitations[J].J Clin Oncol,2000,18:3495-3502.
  • 7Salskov A,Tammisetti VS,Grierson J,et al.FLT:measuring tumar cell proliferation in vivo with positron emission tomography and 3'-deoxy-3'-[^18F]Fluorothymidine[J].Sem Nucl Med,2007,37:429-439.
  • 8Dittmann H,Dohmen BM,Panlsen F,et al.[^18F] FLT PET for diagnosis and staging of thoracic tumor[J].Eur J Nucl Med Mol Imaging,2003,30:1407-1412.
  • 9Chung JK,Kim YK,Kim SK,et al.Usefulness of ^11C-mathionine PET in the evaluation of brain lesions that are hypo-or isometabolic on ^18F-FDG PET[J].Eur J Nucl Med Mol Imaging,2002,29(2):176-182.
  • 10Laveman P,Boerman OC,Corstens FH,et al.Fluorinated amino acids for tumour imaging with positron emission tomography[J].Eur J Nucl Med Mol Imaging,2002,29(5):681-690.

共引文献21

同被引文献21

引证文献4

二级引证文献12