期刊文献+

硒(Ⅳ)的示波极谱测定 被引量:5

OSCILLOPOLAROGRAPHIC DETERMINATION OF SELENIUM(Ⅳ)
下载PDF
导出
摘要 在0.2mol·L-1的甲酸 甲酸钠缓冲液(pH3.5)中,硒(Ⅳ)于-0.70V(vs.SCE)处有一灵敏的极谱波。在1×10-5~0.5g·L-1范围内峰电流与其浓度呈线性关系,检出限为1×10-6g·L-1,加标回收率为95.0%~114.5%,相对标准偏差为4.44%~7.39%。方法适于试样中微量硒的测定。 In a 0.2mol·L^(-1) acetate buffer medium of pH 3.5, selenium(Ⅳ) gives a sensitive polarographic wave at -0.70V (vs. SCE). The peak currents (i_p) keep a linear relationship with the concentrations of Se(Ⅳ) in the range of 1×10^(-5)~0.5g of Se(Ⅳ) perliter. Detection limit of this method attained to a value of 1×10^(-6)g·L^(-1), and recoveries obtained were in the range of 95.0%~114.5%. Precision tests carried out in the concentrations within the working curve gave values of RSD in the range of 4.44%~7.39%. This method has been satisfactorily applied to the analysis of some practical samples containing microamounts of selenium.
作者 李敏 张克荣
出处 《理化检验(化学分册)》 CAS CSCD 北大核心 2004年第3期153-154,共2页 Physical Testing and Chemical Analysis(Part B:Chemical Analysis)
关键词 硒(Ⅳ) 示波极谱法 含量测定 甲酸-甲酸钠缓冲液 Oscillopolarography Selenium(Ⅳ) Water sample Tea sample Standard reference material
  • 相关文献

参考文献2

二级参考文献13

  • 1米瑞华.硒的原子吸收分析进展[J].光谱学与光谱分析,1993,13(4):103-110. 被引量:23
  • 2杨逸萍,厦门大学学报,1987年,26卷,1期,96页
  • 3韩恒斌,分析化学,1985年,13卷,11期,866页
  • 4徐保玲,分析化学,1985年,13卷,29页
  • 5俞穆清,环境化学,1982年,1卷,6期,467页
  • 6Tzeng J,Anal Chim Acta,1978年,101卷,71页
  • 7He Y Z,J Anal Atom Spectrom,1998年,13期,289页
  • 8Liu Yongming,Talanta,1996年,43卷,985页
  • 9Cunna R M,Talanta,1995年,42卷,6期,817页
  • 10Zhang B G,Talanta,1995年,42卷,1095页