摘要 所设计的磁化器场强分布测定仪能够测定磁化器的磁性能参数,由此来确定磁路的最优组合分布.介绍了仪器的工作原理及组成,分析了霍尔传感器的性能和保证测量精度的方法,并给出了实际的测试结果. The instrument designed can determine the magnetic performance of magnetizaers so as to find out the optimal combined distribution of the magnetization. Introduced the theory and composition of the instrument,analyzed the performance of hall sensor,studied the algorithms to insure precision in details and gave the test results.
机构地区 哈尔滨工业大学
出处 《哈尔滨商业大学学报(自然科学版)》 CAS 2004年第1期32-35,共4页 Journal of Harbin University of Commerce:Natural Sciences Edition
关键词 磁场测量 霍尔传感器 计算机控制 智能仪器 磁化器 场强分布 测定仪 油井 measure of magnetizer hall sensor computer control intelligent instrument
  • 相关文献

参考文献6

二级参考文献21

  • 1戴景民,杨茂华,褚载祥.多波长辐射测温仪及其应用[J].红外与毫米波学报,1995,14(6):461-466. 被引量:28
  • 2吴锡令.波导法测量油井持水率的理论、方法和实验研究[J].地球物理学报,1996,39(3):424-430. 被引量:13
  • 3DAI Jingmin, FAN Yi, CHU Zaixiang. Development of a millisecond pulse-heating apparatus in China[J]. International Journal of Thermophysics ,2002,23 : 1401 - 1405.
  • 4CEZAIRLIYAN A. Advances in measurements of thermophysical properties by dynamic techniques [ J ]. High Temp High Press, 1979,11:9 -27.
  • 5CEZAIRLIYAN A. A millisecond-resolution pulse heating system for specific-heat measurements at high temperatures[ A]. MAGLIC K D, CEZAIRLIYAN A, PELETSKII V E Eds. Compendium of Thermophysical Property Measurement Methods. New York:Plenum,1992.
  • 6RIGHINI F, ROSSO A, RUFFINO G. System for fast hagh-temperature measurements [ J ]. High Temp High Press. 1972.4. 597-603.
  • 7CEZAIRLIYAN A, KRISHNAN S, MCCLURE J L. Simultaneous measurements of normal spectral emissivity by spectral radiometry and laser polarimetry at high temperatures in millisecond-resolution pulse-heating experiments:application to molybdenum and tungsten[J]. Int J Thermophysics, 1996, 17 : 1455 - 1473.
  • 8RIGHINI F, SPISIAK J, BUSSOLINO G C. Normal spectral emissivity of niobium ( at 900nm ) by a pulseheating reflectometric technique [ J ]. Int J Thermophysits, 1999, 20:1095 - 1106.
  • 9CEZAIRLIYAN A, KRISHNAN S, BASAK D, et al.Application of laser polarimetry to the measurement of the specific heat capacity of molybdenum in the range 2000 to 2800K by a pulse-heating technique [ J ]. Int J Thermophysics, 1998, 19 : 1267 - 1276.
  • 10RIGHINI F, SPISIAK J, BUSSOLINO G C, et al. Thermophysical properties by a pulse-heating reflectometric technique: niobium, 1100 to 2700K. Int J Thermophysics, 1999, 20:1107 - 1116.