摘要
在脉冲激光激发荧光的研究中由于入射激光强度的起伏会导致所激发荧光的强度随之波动 ,从而给荧光光谱及其待测物质测量结果带来较大误差。采用双光路法 (即在采集荧光谱的同时也采集激发激光强度 ,利用荧光强度与激发光强度的线性相关 ,对实验数据进行相关分析 )可以有效地克服激发光波动对荧光光谱的影响 ,显著降低了测量误差。用此法对若丹明 6G和十二烷基硫酸纳 (SDS)对R6G的荧光增强进行测量 ,可得到相关系数大于 0 9的测量结果。在激发光波动较大时双光路法与单光路采样平均法相比有明显优越性。
The incident pulse intensity wave will cause the induced fluorescence wave, which will lead to the obvious error of experiment results. A double-light-beam correlation method between the incident laser and the fluorescence was used to decrease such error efficiently in the experiments. A good correlation factor over 0.9 was obtained by the method in the experiment of R6G fluorescence enhanced by SDS. The method is better than sampling-averaging method when the incident laser intensity has more waves.