目的 观察尿液十项分析仪 (简称尿仪 )与显微镜检测 (简称镜检 )尿中红细胞的相关性。方法 住院患者送检的晨尿标本 1 4 2 1份 ,同时进行两种方法检测 ,将结果分析比较。结果 两种方法的总符合率为 85 4 % ,其中阴性完全符合率为 1 0 0 % ,阳性完全符合率为 1 4 8%。结论 尿仪检测的的敏感性明显高于镜检 ,尿仪检测BLD≤ 1 +,可忽略镜检 ;>1 +则必须镜检 ;明显不符时 ,应结合临床综合判断 。
AIM To observe the correlation of urine ten analysis apparatuses (abbreviation urine apparatus) and microscopic examination for middle red blood cell of urine. METHODS The patient being in hospital send morning checked up 1 421 of the specimen in the urine, and carries on at the same time two kinds of methods testing, and compares the interpretation of result. RESULTS The complete rate fitting of the among them negative completely rate fitting of the rate fitting always of two kinds of methods for 85 4%, for 100%, s masculine is 14 8%. CONCLUSIONS Urine apparatus the person who tests clear more than looking glass of sensitivity is checked up. Urine apparatus checkout BLD ≤ 1+, can neglect the looking glass checking up But >1+ must be checked up by the looking glass Should combine the clinical synthetical judgement when clearly not agreeing with, and the doctor keeps the contact when necessary.