期刊文献+

GFAAS测定钼时两种石墨管的比较研究

Comparison of Two Kinds of Graphite Tubes in Determination of Molybdenum by GFAAS
下载PDF
导出
摘要 本文研究比较了全热解石墨管与热解涂层石墨管在未配备温度控制装置的石墨炉原子吸收分光光度计上测定高温难熔元素钼时的测量性能。结果表明,全热解石墨管优于热解涂层石墨管。本文对两种石墨管性能差异的主要原因进行了初步探讨。 Differences between total pyrolytic graphite tubes and pyrolytically coated graphite tubes at characteris-tis in determination of molybdenum by graphite furnace atomic absorption spectrometry (GFAAS) havebeen investigated.The atomic absorption spetrometer used in this paper has not been equipped-with the tem-perature accessory.The results have shown that the total pyrolytic tubes are superior to the pyrolyticallycoated graphite tubes.
出处 《材料工程》 EI CAS CSCD 北大核心 1993年第4期33-35,共3页 Journal of Materials Engineering
基金 国家自然科学基金
关键词 全热解石墨管 热解涂层 石墨管 Absorption spectroscopy Graphite Spectroscopic analysis Tubes (components)
  • 相关文献

参考文献2

  • 1周南根,W. Frech,L. De Galan.论电热原子吸收光谱分析中原子化加热速率与峰值的关系[J]光谱学与光谱分析,1987(02).
  • 2马怡载,李绍元,张仲明,吴正科,冯皙,苏雯,孙涤君.石墨炉原子吸收分析中升温速率对灵敏度影响的研究[J]化学通报,1984(08).