期刊文献+

升温速率对氧化石墨烯热重分析法表征研究的影响

The Impact of Heating Rate on the Characterization of Graphene Oxide by Thermogravimetric Analysis
下载PDF
导出
摘要 本研究探讨了升温速率对氧化石墨烯热重分析(TGA)的影响。实验结果表明,在5℃/min和7℃/min的升温速率下,氧化石墨烯的TGA测试结果稳定,未发生样品溢出,且升温速率对测试结果的影响不显著。然而,当升温速率增至8℃/min时,出现了样品溢出现象。因此,建议在室温至300℃的区间内,采用2℃/min至7℃/min的升温速率,其中5℃/min为推荐值。在300℃以上,升温速率可以提高至10℃/min至15℃/min,直至1000℃。这种分段升温策略有效防止了样品溢出,确保了测试结果的准确性,并为氧化石墨烯的热稳定性评估提供了可靠的数据支持。This study aims to investigate the impact of heating rates on the thermogravimetric analysis (TGA) of graphene oxide (GO). The experiments indicated that when the heating rate was set to 5˚C/min and 7˚C /min, the TGA process of GO was stable, with no sample overflow observed, and the effect of different heating rates on the results was negligible. However, at a heating rate of 8˚C /min, sample overflow occurred. Therefore, it is recommended to use a constant heating rate of 2˚C/min to 7˚C/min for the phase from room temperature to 300˚C, with 5˚C/min being the preferred choice. Beyond 300˚C, the heating rate can be increased to 10˚C/min to 15˚C/min, up to 1000˚C. This segmented heating strategy effectively prevents sample overflow, ensures the accuracy of the test, and provides reliable data support for the assessment of the thermal stability of graphene oxide.
出处 《材料科学》 2024年第11期1549-1553,共5页 Material Sciences