期刊文献+

Mechanism of Cuff-Less Blood Pressure Measurement Using MMSB

Mechanism of Cuff-Less Blood Pressure Measurement Using MMSB
下载PDF
导出
摘要 Continuous measurement of blood pressure based on pulse transit time (PTT) using GMR sensors is the state-of-art non-invasive cuffless method in which modulated magnetic signature of blood (MMSB) is used. In this paper, the mechanism of MMSB is investigated. According to the experimental results, it is found that both blood pulse flowing through the applied magnetic field and the displacement of the GMR sensor caused by blood pulse contribute to the disturbance of magnetic field detected by GMR sensors. The feasibility of MMSB method is discussed as well. Continuous measurement of blood pressure based on pulse transit time (PTT) using GMR sensors is the state-of-art non-invasive cuffless method in which modulated magnetic signature of blood (MMSB) is used. In this paper, the mechanism of MMSB is investigated. According to the experimental results, it is found that both blood pulse flowing through the applied magnetic field and the displacement of the GMR sensor caused by blood pulse contribute to the disturbance of magnetic field detected by GMR sensors. The feasibility of MMSB method is discussed as well.
出处 《Engineering(科研)》 2013年第10期123-125,共3页 工程(英文)(1947-3931)
关键词 BLOOD PRESSURE Magnetic Sensor PULSE TRANSIT Time Blood Pressure Magnetic Sensor Pulse Transit Time