Accuracy of Medical Oxygen Flowmeters: A Multicentric Field Study

Accuracy of Medical Oxygen Flowmeters: A Multicentric Field Study
下载PDF
导出
摘要 The accuracy of 476 oxygen flowmeters was investigated using a thermal mass flowmeter in eight hospitals in France and Belgium. Different oxygen flow rates (2 to 15 l/min) were evaluated at the patient’s bed. When the sample was considered as a whole, the accuracy of delivered flow was acceptable but precision was poor. The variability of the delivered flow between devices was greater when a low flow rate was required. Compensated-pressure oxygen flowmeters for these low rates were more accurate than their non-compensated counterparts. This study emphasizes the need to individually adapt the oxygen flow rate each time a patient has to move from one flowmeter to another. The accuracy of 476 oxygen flowmeters was investigated using a thermal mass flowmeter in eight hospitals in France and Belgium. Different oxygen flow rates (2 to 15 l/min) were evaluated at the patient’s bed. When the sample was considered as a whole, the accuracy of delivered flow was acceptable but precision was poor. The variability of the delivered flow between devices was greater when a low flow rate was required. Compensated-pressure oxygen flowmeters for these low rates were more accurate than their non-compensated counterparts. This study emphasizes the need to individually adapt the oxygen flow rate each time a patient has to move from one flowmeter to another.
出处 《Health》 2014年第15期1978-1983,共6页 健康(英文)
关键词 MEDICAL Device ACCURACY OXYGEN Therapy FLOWMETER Medical Device Accuracy Oxygen Therapy Flowmeter
  • 相关文献

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部