Dosimetric Characteristics of 6 MV Medical Linac at BAEC

Dosimetric Characteristics of 6 MV Medical Linac at BAEC
下载PDF
导出
摘要 Dosimetric characteristic is one of the essential parameters of a medical linear accelerator (LINAC), which must be obtained before clinical use. The dosimetric characteristics for 6 MV photon beam were measured and compared with the corresponding published data. The study was done using a Varian linear accelerator (Model Clinac-iX) at the Institute of Nuclear Medical Physics (INMP), AERE, Savar, Dhaka, under the Bangladesh Atomic Energy Commission (BAEC). The data is taken for 10 field sizes (2 × 2, 3 × 3, 5 × 5, 7 × 7, 10 × 10, 15 × 15, 20 × 20, 25 × 25, 30 × 30 and 40 × 40 cm<sup>2</sup>) at same conditions. The measured Percent Depth Dose (PDD) curves were obtained for 6 MV photon beams with the field as mentioned above and compared with the calculated PDD curves. The measured depth dose (D<sub>max</sub>) for reference field size (FS) 10 × 10 cm<sup>2</sup> is 15.99 mm, and the PDD at 10 cm depth (D<sub>10</sub>) is 66.87% for 6 MV photon energies that are found to be compatible with the published report BJR supplement 25. The measured PDD curves for photon energies show a good agreement with the standard PDD curves. The photon beam dosimetry data found in the current study are compatible and all the tolerances are within the clinically acceptable tolerance limit. Dosimetric characteristic is one of the essential parameters of a medical linear accelerator (LINAC), which must be obtained before clinical use. The dosimetric characteristics for 6 MV photon beam were measured and compared with the corresponding published data. The study was done using a Varian linear accelerator (Model Clinac-iX) at the Institute of Nuclear Medical Physics (INMP), AERE, Savar, Dhaka, under the Bangladesh Atomic Energy Commission (BAEC). The data is taken for 10 field sizes (2 × 2, 3 × 3, 5 × 5, 7 × 7, 10 × 10, 15 × 15, 20 × 20, 25 × 25, 30 × 30 and 40 × 40 cm<sup>2</sup>) at same conditions. The measured Percent Depth Dose (PDD) curves were obtained for 6 MV photon beams with the field as mentioned above and compared with the calculated PDD curves. The measured depth dose (D<sub>max</sub>) for reference field size (FS) 10 × 10 cm<sup>2</sup> is 15.99 mm, and the PDD at 10 cm depth (D<sub>10</sub>) is 66.87% for 6 MV photon energies that are found to be compatible with the published report BJR supplement 25. The measured PDD curves for photon energies show a good agreement with the standard PDD curves. The photon beam dosimetry data found in the current study are compatible and all the tolerances are within the clinically acceptable tolerance limit.
作者 Sudeb Kumar Roy Pretam Kumar Das Rajada Khatun Md. Ashikur Rahman Shirin Akter Tushar Kumar Mohammad Monjur Ahasan Sudeb Kumar Roy;Pretam Kumar Das;Rajada Khatun;Md. Ashikur Rahman;Shirin Akter;Tushar Kumar;Mohammad Monjur Ahasan(Department of Physics, Pabna University of Science and Technology, Pabna, Bangladesh;Medical Physics Division, Atomic Energy Centre, Dhaka, Bangladesh;Institutes of Nuclear Medical Physics, Atomic Energy Research Establishment, Savar, Dhaka, Bangladesh)
出处 《International Journal of Medical Physics, Clinical Engineering and Radiation Oncology》 2021年第1期38-46,共9页 医学物理学、临床工程、放射肿瘤学(英文)
关键词 Linac Machines 3D Water Phantom Ionization Chamber ELECTROMETER Linac Machines 3D Water Phantom Ionization Chamber Electrometer
  • 相关文献

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部