Computed Tomography Protocol Optimisation for Pediatric Head Trauma: Radiation Dose and Image Quality Assessment

Computed Tomography Protocol Optimisation for Pediatric Head Trauma: Radiation Dose and Image Quality Assessment
下载PDF
导出
摘要 Purpose: Children are sometimes examined with Computed Tomography protocols designed for adults, leading to radiation doses higher than necessary. Lack of optimisation could lead to image quality higher than what is needed for diagnostic purposes with associated high doses to patients. Optimising the protocols for paediatric head trauma CT imaging will reduce radiation dose. Objective: The study aimed to optimise radiation dose and assess the image quality for a set of protocols by evaluating noise, a contrast to noise ratio, modulation transfer function and noise power spectrum. Methods: Somaton Sensation 64 was used to scan the head of an anthropomorphic phantom with a set of protocols. ImageJ software was used to analyse the paediatric head image from the scanner. IMPACTSCAN dosimeter software was used to evaluate the radiation dose to the various organs in the head. MATLAB was used to analyse the Modulation Transfer Function and the Noise Power. Results: The estimated Computed Tomography Dose Index volume (CTDI<sub>vol</sub>) increased with increasing tube current and tube voltage. The high pitch of 0.9 gave a lower dose than the 0.5 pitch. The eye lens received the highest radiation dose (39.2 mGy) whiles the thyroid received the least radiation dose (13.7 mGy). There was an increase in noise (62.46) when the H60 kernel was used and a lower noise (8.829) was noticed when the H30 kernel was used. Conclusion: The results obtained show that the H30 kernel (smooth kernel) gave higher values for noise and contrast to noise ratio (CNR) than the H60 kernel (sharp kernel). The H60 kernel produced high values for the modulation transfer function (MTF) and noise power spectrum (NPS). The eye lens received the highest radiation dose. Purpose: Children are sometimes examined with Computed Tomography protocols designed for adults, leading to radiation doses higher than necessary. Lack of optimisation could lead to image quality higher than what is needed for diagnostic purposes with associated high doses to patients. Optimising the protocols for paediatric head trauma CT imaging will reduce radiation dose. Objective: The study aimed to optimise radiation dose and assess the image quality for a set of protocols by evaluating noise, a contrast to noise ratio, modulation transfer function and noise power spectrum. Methods: Somaton Sensation 64 was used to scan the head of an anthropomorphic phantom with a set of protocols. ImageJ software was used to analyse the paediatric head image from the scanner. IMPACTSCAN dosimeter software was used to evaluate the radiation dose to the various organs in the head. MATLAB was used to analyse the Modulation Transfer Function and the Noise Power. Results: The estimated Computed Tomography Dose Index volume (CTDI<sub>vol</sub>) increased with increasing tube current and tube voltage. The high pitch of 0.9 gave a lower dose than the 0.5 pitch. The eye lens received the highest radiation dose (39.2 mGy) whiles the thyroid received the least radiation dose (13.7 mGy). There was an increase in noise (62.46) when the H60 kernel was used and a lower noise (8.829) was noticed when the H30 kernel was used. Conclusion: The results obtained show that the H30 kernel (smooth kernel) gave higher values for noise and contrast to noise ratio (CNR) than the H60 kernel (sharp kernel). The H60 kernel produced high values for the modulation transfer function (MTF) and noise power spectrum (NPS). The eye lens received the highest radiation dose.
作者 Evelyn Anaafi Mary Boadu Albertina Rusandu Mercy Afadzi Kwame Anokye Amoabeng Evelyn Anaafi;Mary Boadu;Albertina Rusandu;Mercy Afadzi;Kwame Anokye Amoabeng(Department of Medical Physics, Graduate School of Nuclear and Allied Sciences, University of Ghana, Legon;Sweden-Ghana Medical Center, Accra, Ghana;Department of Circulation and Medical Imaging, Norwegian University of Science and Technology (NTNU), Trondheim, Norway;Department of Physics, Norwegian University of Science and Technology (NTNU), Trondheim, Norway;National Centre for Radiotherapy and Nuclear Medicine, Korlebu Teaching Hospital, Accra, Ghana)
出处 《International Journal of Medical Physics, Clinical Engineering and Radiation Oncology》 2022年第3期160-175,共16页 医学物理学、临床工程、放射肿瘤学(英文)
关键词 Image Quality Radiation Dose Modulation Transfer Function Noise Power Spectrum OPTIMIZATION Image Quality Radiation Dose Modulation Transfer Function Noise Power Spectrum Optimization
  • 相关文献

参考文献2

共引文献6

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部