Reduction of the Radiation Dose by Decreasing the Tube Current without Degradation of Low-Contrast Detectability on Abdominal Multi-Detector Row CT: A Phantom-Based Study

Reduction of the Radiation Dose by Decreasing the Tube Current without Degradation of Low-Contrast Detectability on Abdominal Multi-Detector Row CT: A Phantom-Based Study
下载PDF
导出
摘要 As use of the lowest acceptable radiation dose during routine diagnostic imaging is important, we determined the optimal tube current without degradation of low-contrast detectability on abdominal multi-detector row CT (MDCT). CT scanning was performed with a Catphan®?500 phantom. The optimal tube current was 300 mA on 64-MDCT and 160 mA on 8-MDCT, with a fixed voltage of 120 kV. Reduction of the radiation dose in abdominal CT scanning by lowering the tube current proved to be feasible. As use of the lowest acceptable radiation dose during routine diagnostic imaging is important, we determined the optimal tube current without degradation of low-contrast detectability on abdominal multi-detector row CT (MDCT). CT scanning was performed with a Catphan®?500 phantom. The optimal tube current was 300 mA on 64-MDCT and 160 mA on 8-MDCT, with a fixed voltage of 120 kV. Reduction of the radiation dose in abdominal CT scanning by lowering the tube current proved to be feasible.
出处 《Open Journal of Medical Imaging》 2013年第4期110-115,共6页 医学影像期刊(英文)
关键词 Tube Current Radiation DOSE COMPUTED Tomography X-Ray ABDOMINAL Imaging Tube Current Radiation Dose Computed Tomography X-Ray Abdominal Imaging