期刊文献+

Evaluation of titanium dental implants after early failure of osseointegration by means of X-ray photoelectron spectoscopy, electron microscopy and histological studies

Evaluation of titanium dental implants after early failure of osseointegration by means of X-ray photoelectron spectoscopy, electron microscopy and histological studies
下载PDF
导出
摘要 In this work, we analysed 56 clinically failed and retrieved implants by means of scanning electron microscopy (SEM), X-ray photoelectron spectroscopy (XPS) and histological studies. The surface contamination was compared to that of unused control implants and with that of the same implants after cleaning in a basic medium. The surfaces of the unused implants presented considerable contamination. In particular, high levels of carbon were detected. The nature of this C was elucidated by XPS analysis of the lubricant used in the machining process. The same contamination was observed in the retrieved implants. Histological studies were carried out by means of light microscopy. Fibrosis and granulomatous lesions were detected in the tissues. XPS analysis indicated the presence of traces of other elements (Na, Ca, Zn, S, F, etc.) that were not related to impurities in cpTi. We examined a cleaning process in a basic medium that eliminates the organic components of the implant surfaces. The cleaned implants were implanted in the patients and the results were excellent. None of the implants failed in following 7 months. In this work, we analysed 56 clinically failed and retrieved implants by means of scanning electron microscopy (SEM), X-ray photoelectron spectroscopy (XPS) and histological studies. The surface contamination was compared to that of unused control implants and with that of the same implants after cleaning in a basic medium. The surfaces of the unused implants presented considerable contamination. In particular, high levels of carbon were detected. The nature of this C was elucidated by XPS analysis of the lubricant used in the machining process. The same contamination was observed in the retrieved implants. Histological studies were carried out by means of light microscopy. Fibrosis and granulomatous lesions were detected in the tissues. XPS analysis indicated the presence of traces of other elements (Na, Ca, Zn, S, F, etc.) that were not related to impurities in cpTi. We examined a cleaning process in a basic medium that eliminates the organic components of the implant surfaces. The cleaned implants were implanted in the patients and the results were excellent. None of the implants failed in following 7 months.
机构地区 不详
出处 《Journal of Biomedical Science and Engineering》 2010年第11期1073-1077,共5页 生物医学工程(英文)
关键词 DENTAL IMPLANT TITANIUM CONTAMINATION OSSEOINTEGRATION Dental Implant Titanium Contamination Osseointegration
  • 相关文献

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部