Determination of Residual Acrylamide in Medical Polyacrylamide Hydrogel by High Performance Liquid Chromatography tandem Mass Spectroscopy 被引量:3
1
作者 WEI-WEI LI HUI LI +1 位作者 ZHI-FEI LIU qun qiao 《Biomedical and Environmental Sciences》 SCIE CAS CSCD 2009年第1期28-31,共4页
Objective To determine residual acrylamide in medical polyacrylamide hydrogel by high performance liquid chromatography tandem mass spectroscopy (HPLC-MS). Methods After 13C3 labeled acrylamide was added, the sample... Objective To determine residual acrylamide in medical polyacrylamide hydrogel by high performance liquid chromatography tandem mass spectroscopy (HPLC-MS). Methods After 13C3 labeled acrylamide was added, the sample was extracted with water and then cleaned up with ExtrelutTM 20. The polyacrylamide hydrogel sample and 20 clinical cases were analyzed by HPLC-MS/MS and isotope dilution quantifying technique in selected reaction monitoring (SRM) mode. Results Acrylamide was separated from polyacrylamide hydrogel. The concentration of acrylamide in polyacrylamide hydrogel ranged from 3.9×10^-9 to 3.1×10^- 8g/L in the 20 clinical cases. The peak area was favorable linear and the range was up to 3 000 μg/L. The recovery rate was 103.1% with a relative standard deviation (RSD) of 6.20%, when the mark level was 50 lag/L. Conelusion HPLC-MS is a rapid, accurate, and sensitive method for the determination of residual acrylamide in medical polyacrylamide hydrogel. 展开更多
关键词 High performance liquid chromatography-mass spectroscopy Polyacrylamide hydrogel ACRYLAMIDE