Two rapid, sensitive and reliable immunoassay methods, namely competitive indirect enzyme-linked immunosorbent assay (CI-ELISA) and colloidal gold-based immunochromatographic assay (CGIA), were developed to detect ofl...Two rapid, sensitive and reliable immunoassay methods, namely competitive indirect enzyme-linked immunosorbent assay (CI-ELISA) and colloidal gold-based immunochromatographic assay (CGIA), were developed to detect ofloxacin (OFL). The linear range of the CI-ELISA was from 0.5 to 128 ng/mL with a limit of detection (LOD) of 0.35 ng/mL. Good recoveries were obtained in analyzing simulated swine urine samples. The CGIA could accurately estimate OFL at concentrations as low as 10 ng/mL in less than 10 min, and test results were read visually without any instrument.展开更多
文摘Two rapid, sensitive and reliable immunoassay methods, namely competitive indirect enzyme-linked immunosorbent assay (CI-ELISA) and colloidal gold-based immunochromatographic assay (CGIA), were developed to detect ofloxacin (OFL). The linear range of the CI-ELISA was from 0.5 to 128 ng/mL with a limit of detection (LOD) of 0.35 ng/mL. Good recoveries were obtained in analyzing simulated swine urine samples. The CGIA could accurately estimate OFL at concentrations as low as 10 ng/mL in less than 10 min, and test results were read visually without any instrument.