Glaucoma surgery experiments using digital microscope-integrated optical coherence tomography and OCT-compatible instruments 被引量:1
1
作者 Tianliang Jiang Jinyu Fan +2 位作者 Ning Tang Yi He Guohua Shi 《Journal of Innovative Optical Health Sciences》 SCIE EI CSCD 2024年第5期105-113,共9页
There is a certain failure rate in traditional glaucoma surgery because of the lack of depth information in microscope images.In this work,we present a digital microscope-integrated optical coherence tomography(MIOCT)... There is a certain failure rate in traditional glaucoma surgery because of the lack of depth information in microscope images.In this work,we present a digital microscope-integrated optical coherence tomography(MIOCT)system and several custom-made OCT-compatible instruments for glaucoma surgery.Sixteen ophthalmologists were asked to perform trabeculectomy and canaloplasty on live porcine eyes using the system and instruments.After surgery,a subjective feedback survey about the user experience was taken.The experiment results showed that our system can help surgeons easily locate important tissue structures during surgery.The custom-made instruments also solved the shadowing problem in OCT imaging.Surgeons preferred to use the system in their future practice. 展开更多
关键词 microscope-integrated optical coherence tomography OCT-compatible instruments glaucoma surgery TRABECULECTOMY CANALOPLASTY