A high-sensitivity DC magneto-optical Kerr effect(MOKE)apparatus is described in this Letter.Via detailed analysis on several dominatingnoise sources,we have proposed solutions that significantly lower the MOKE noise,...A high-sensitivity DC magneto-optical Kerr effect(MOKE)apparatus is described in this Letter.Via detailed analysis on several dominatingnoise sources,we have proposed solutions that significantly lower the MOKE noise,and a sensitivity of 1.5×10^(-7)rad/√Hz is achieved with long-term stability.The sensitivity of the apparatus is tested by measuring a wedgeshaped Ni thin film on SiO_(2)with Ni thickness varying from 0 to 3 nm.A noise floor of 1.5×10^(-8)rad is demonstrated.The possibility of further improving sensitivity to 10^(-9)rad via applying AC modulation is also discussed.展开更多
基金support from the National Natural Science Foundation of China(Nos.12125403 and 11874123)the Shanghai Science and Technology Committee(No.20ZR1406000)。
文摘A high-sensitivity DC magneto-optical Kerr effect(MOKE)apparatus is described in this Letter.Via detailed analysis on several dominatingnoise sources,we have proposed solutions that significantly lower the MOKE noise,and a sensitivity of 1.5×10^(-7)rad/√Hz is achieved with long-term stability.The sensitivity of the apparatus is tested by measuring a wedgeshaped Ni thin film on SiO_(2)with Ni thickness varying from 0 to 3 nm.A noise floor of 1.5×10^(-8)rad is demonstrated.The possibility of further improving sensitivity to 10^(-9)rad via applying AC modulation is also discussed.