The caesium atomic fountain clock is a primary frequency standard.During its operation,a Majorana transition frequency shift will occur once a magnetic field at some special locations along the atomic trajectory is si...The caesium atomic fountain clock is a primary frequency standard.During its operation,a Majorana transition frequency shift will occur once a magnetic field at some special locations along the atomic trajectory is singular.In this study,by developing a physical model,we analyzed the magnetic field requirements for atomic adiabatic transition and calculated the influence of the Majorana atomic transition on the atomic state via a quantum method.Based on the simulation results for the magnetic field in the fountain clock,we applied the Monte Carlo method to simulate the relationship between the Majorana transition frequency shift and the magnetic field at the entrance of the magnetic shielding,as well as the initial atomic population.Measurement of the Majorana transition frequency shift was realized by state-selecting asymmetrically populated atoms.The relationship between the Majorana transition frequency shift and the axial magnetic field at the entrance of the magnetic shielding was obtained.The measured results were essentially consistent with the calculated results.Thus,the magnetic field at the entrance of the magnetic shielding was configured,and the Majorana transition frequency shift of the fountain clock was calculated to be 4.57×10^(-18).展开更多
Caesium atomic fountain clock is a primary frequency standard,which realizes the duration of second.Its performance is mostly dominated by the frequency accuracy,and the C-field induced second-order Zeeman frequency s...Caesium atomic fountain clock is a primary frequency standard,which realizes the duration of second.Its performance is mostly dominated by the frequency accuracy,and the C-field induced second-order Zeeman frequency shift is the major effect,which limits the accuracy improvement.By applying a high-precision current supply and high-performance magnetic shieldings,the C-field stability has been improved significantly.In order to achieve a uniform C-field,this paper proposes a doubly wound C-field solenoid,which compensates the radial magnetic field along the atomic flight region generated by the lead-out single wire and improves the accuracy evaluation of second-order Zeeman frequency shift.Based on the stable and uniform C-field,we launch the selected atoms to different heights and record the magnetically sensitive Ramsey transition|F=3,mF=-1→|F=4,mF=-1 central frequency,obtaining this frequency shift as 131.03×10^(-15)and constructing the C-field profile(σ=0.15 n T).Meanwhile,during normal operation,we lock NTSC-F2 to the central frequency of the magnetically sensitive Ramsey transition|F=3,mF=-1→|F=4,mF=-1 fringe for ten consecutive days and record this frequency fluctuation in time domain.The first evaluation of second-order Zeeman frequency shift uncertainty is 0.10×10^(-15).The total deviation of the frequency fluctuation on the clock transition induced by the C-field instability is less than 2.6×10^(-17).Compared with NTSC-F1,NTSC-F2,there appears a significant improvement.展开更多
基金Project supported by the National Natural Science Foundation of China(Grant No.12173044)Research and Development Project of Scientific Research Instruments and Equipment of Chinese Academy of Sciences(Grant No.YJKYYQ20200020)+1 种基金Large Research Infrastructures Improvement Funds of Chinese Academy of Sciences(Grant No.DSS-WXGZ-2020-0005)Chinese Academy of Sciences for Western Young Scholars(Grant Nos.XAB2018A06,XAB2019A07,and XAB2018B16)。
文摘The caesium atomic fountain clock is a primary frequency standard.During its operation,a Majorana transition frequency shift will occur once a magnetic field at some special locations along the atomic trajectory is singular.In this study,by developing a physical model,we analyzed the magnetic field requirements for atomic adiabatic transition and calculated the influence of the Majorana atomic transition on the atomic state via a quantum method.Based on the simulation results for the magnetic field in the fountain clock,we applied the Monte Carlo method to simulate the relationship between the Majorana transition frequency shift and the magnetic field at the entrance of the magnetic shielding,as well as the initial atomic population.Measurement of the Majorana transition frequency shift was realized by state-selecting asymmetrically populated atoms.The relationship between the Majorana transition frequency shift and the axial magnetic field at the entrance of the magnetic shielding was obtained.The measured results were essentially consistent with the calculated results.Thus,the magnetic field at the entrance of the magnetic shielding was configured,and the Majorana transition frequency shift of the fountain clock was calculated to be 4.57×10^(-18).
基金the National Key R&D Program of China(Grant No.2016YFF0200202)the Maintenance and Reformation Program for the Major Science and Technology Fundamental Devices of the Chinese Academy of Sciences(Grant No.DSS-WXGZ-2020-0005)the Foundation for Western Young Scholars,China(Grant No.XAB2018A06)。
文摘Caesium atomic fountain clock is a primary frequency standard,which realizes the duration of second.Its performance is mostly dominated by the frequency accuracy,and the C-field induced second-order Zeeman frequency shift is the major effect,which limits the accuracy improvement.By applying a high-precision current supply and high-performance magnetic shieldings,the C-field stability has been improved significantly.In order to achieve a uniform C-field,this paper proposes a doubly wound C-field solenoid,which compensates the radial magnetic field along the atomic flight region generated by the lead-out single wire and improves the accuracy evaluation of second-order Zeeman frequency shift.Based on the stable and uniform C-field,we launch the selected atoms to different heights and record the magnetically sensitive Ramsey transition|F=3,mF=-1→|F=4,mF=-1 central frequency,obtaining this frequency shift as 131.03×10^(-15)and constructing the C-field profile(σ=0.15 n T).Meanwhile,during normal operation,we lock NTSC-F2 to the central frequency of the magnetically sensitive Ramsey transition|F=3,mF=-1→|F=4,mF=-1 fringe for ten consecutive days and record this frequency fluctuation in time domain.The first evaluation of second-order Zeeman frequency shift uncertainty is 0.10×10^(-15).The total deviation of the frequency fluctuation on the clock transition induced by the C-field instability is less than 2.6×10^(-17).Compared with NTSC-F1,NTSC-F2,there appears a significant improvement.