Modulation domain measurement can be u sed to analyze the frequency,phase,and time intervals which transiently change with time.During the implementation of modulation domain analyzer,meas urement errors can be introd...Modulation domain measurement can be u sed to analyze the frequency,phase,and time intervals which transiently change with time.During the implementation of modulation domain analyzer,meas urement errors can be introduced in multiple parts.Aiming at a typical implementation of modul ation domain analyzer,this paper analyzes the inherent error between channe l s,temperature drift error in time-to-digital converter(TDC),random err or,time-base error and their influence on the modulation domain measurement.It also gives a correction sche me and comparison of the measurement results.The frequency resolution can be im proved from 10 to 12 bit/s after correction.展开更多
文摘Modulation domain measurement can be u sed to analyze the frequency,phase,and time intervals which transiently change with time.During the implementation of modulation domain analyzer,meas urement errors can be introduced in multiple parts.Aiming at a typical implementation of modul ation domain analyzer,this paper analyzes the inherent error between channe l s,temperature drift error in time-to-digital converter(TDC),random err or,time-base error and their influence on the modulation domain measurement.It also gives a correction sche me and comparison of the measurement results.The frequency resolution can be im proved from 10 to 12 bit/s after correction.