期刊文献+
共找到1篇文章
< 1 >
每页显示 20 50 100
Measurements of anterior segment parameters using three different non-contact optical devices in keratoconus patients
1
作者 ahmet taylan yazici Gkhan Pekel +4 位作者 Ercüment Bozkurt Yusuf Yildirim Evre Pekel ahmet Demirok mer Faruk Yilmaz 《International Journal of Ophthalmology(English edition)》 SCIE CAS 2013年第4期521-525,共5页
AIM:To evaluate the measurements of anterior segment parameters using three different non-contact optical devices in keratoconus patients.·METHODS:A hundred and one eyes of 55 keratoconus patients were enrolled i... AIM:To evaluate the measurements of anterior segment parameters using three different non-contact optical devices in keratoconus patients.·METHODS:A hundred and one eyes of 55 keratoconus patients were enrolled in this study.The mean age was26.2 ±8.9 years.The inclusion criteria were keratoconus stage I to III according to the Amsler-Krumeich keratoconus classification.All the measurements were done by the same operator,under the mesopic light condition and repeated with three different optical methods;Visante,Orbscan and Pentacam.The evaluated anterior segment parameters were anterior chamber depth(ACD),central and thinnest corneal thickness(CCT and TCT) and pupil diameter(PD).·RESULTS:The mean CCT measured by Visante,Orbscan and Pentacam were as follows:462.0 ±48.1μm,463.9±60.9μm,476.5±45.3μm,respectively(P =0.873).The mean ACD values were 3.34±0.33mm,3.26±0.33mm,3.49±0.40mm,respectively(P=0.118).The mean PD measurements were 5.11 ±1.14mm,4.80 ±0.85mm,3.80 ±1.38mm,respectively(P 【0.001).The mean TCT measurements of Visante,Orbscan and Pentacam were 437.9 ±48.2μm,447.6±60.6μm and 459.9±44.0μm,respectively(P =0.214).The Visante and Orbscan measured CCT similarly,while Pentacam measured CCT thicker than the other two.The Visante measured TCT thinner than the other two devices.In ACD measurements,Orbscan was the one giving the lowest values.PD was measured differently by the devices.·CONCLUSION:Although TCT,CCT and ACD measurements acquired by Visante,Orbscan and Pentacam in keratoconus patients are similar,PDmeasurements show large differences among the devices. 展开更多
关键词 Visante ORBSCAN PENTACAM corneal pachymetry anterior chamber depth pupil diameter
下载PDF
上一页 1 下一页 到第
使用帮助 返回顶部