作者 Dawie van den Heever Cornie Scheffer +1 位作者 Pieter Erasmus edwin dillon 《Journal of Biomedical Science and Engineering》 2012年第12期729-736,共8页
It is suggested that unicompartmental knee replacement (UKR) offers the potential to restore normal knee kinematics better than total knee replacement (TKR) because of retaining the cruciate ligaments, and better pres... It is suggested that unicompartmental knee replacement (UKR) offers the potential to restore normal knee kinematics better than total knee replacement (TKR) because of retaining the cruciate ligaments, and better preservation of the overall geometry. It was hypothesized that patient-specific UKR would restore normal knee kinematics even better because of a customised articular shape. A comparative kinematics study was conducted on three cadaver limbs using two different test setups, a loaded ankle rig and an unloaded ankle rig. Kinematics was compared between a patient-specific UKR and a conventional fixed-bearing UKR. Both the UKRs showed similar kinematic patterns to the normal knee using both the test apparatus. The patient-specific UKR showed good results and with the other benefits it shows potential to dramatically improve clinical outcomes of knee replacement surgery. 展开更多