A multianalyte biosensor for the simultaneous determination of glucose and galactose was developed by immobilizing glucose oxidase (GOD) and galactose oxidase (GAO) on Nafion-modified thin film platinum disk electrod...A multianalyte biosensor for the simultaneous determination of glucose and galactose was developed by immobilizing glucose oxidase (GOD) and galactose oxidase (GAO) on Nafion-modified thin film platinum disk electrodes. The dual Pt working electrodes with disk shape and the surrounding ring shaped counter electrode were fabricated by thin film technology, which were integrated onto the same microchip. The response of the designed biosensor for glucose and galactose were linear up to 6.0 mmol/L and 3.5 mmol/L with sensitivities of 0.3 mA/mmol/L and 0.12 mA/mmol/L, respectively. No cross-talking effect was observed.展开更多
文摘A multianalyte biosensor for the simultaneous determination of glucose and galactose was developed by immobilizing glucose oxidase (GOD) and galactose oxidase (GAO) on Nafion-modified thin film platinum disk electrodes. The dual Pt working electrodes with disk shape and the surrounding ring shaped counter electrode were fabricated by thin film technology, which were integrated onto the same microchip. The response of the designed biosensor for glucose and galactose were linear up to 6.0 mmol/L and 3.5 mmol/L with sensitivities of 0.3 mA/mmol/L and 0.12 mA/mmol/L, respectively. No cross-talking effect was observed.