AIM:To compare the RTVue spectral optical coherence tomography(SD-OCT),Sirius Scheimpflug-Placido topographer,Lenstar optical low coherence reflectometry(OLCR)and ultrasound pachymetry(USP)devices in terms of their ag...AIM:To compare the RTVue spectral optical coherence tomography(SD-OCT),Sirius Scheimpflug-Placido topographer,Lenstar optical low coherence reflectometry(OLCR)and ultrasound pachymetry(USP)devices in terms of their agreement and repeatability of measuring central corneal thickness(CCT).METHODS:In this prospective study,50 eyes of 50patients were included.Three repeated measures were obtained using SD-OCT,Scheimpflug-Placido topographer and USP and five measurements were determined with the OLCR.Bland-Altman plots were used to assess agreement among the instruments,and 95%limits of agreement(LoA)for each comparison were calculated.Intra-examiner repeatability was assessed using intraclass correlation coefficients(ICCs).RESULTS:The mean CCT by SD-OCT,ScheimpflugPlacido topographer,OLCR,and USP were 525.90±34.08μm,525.92±34.10μm,530.30±35.62μm,and543.50±37.11μm respectively.All 4 modalities of CCT measurements correlated closely with each other,with Pearson correlation coefficients ranging from 0.977 to0.995.The mean differences(and upper/lower LoA)for CCT measurements were-0.05±6.77μm(13.3/-13.3)between SD-OCT and Scheimpflug-Placido topographer,4.38±3.79μm(11.8/-3.1)between OLCR and SD-OCT,4.38±6.03μm(16.2/-7.5)between OLCR and ScheimpflugPlacido topographer,13.20±6.46μm(25.9/0.5)between USP and OLCR,17.59±6.76μm(30.8/4.3)between USP and SD-OCT,and 17.58±8.13μm(33.5/1.6)between USP and Scheimpflug-Placido topographer.Intra-examiner repeatability was excellent for all devices with ICCs】0.98.CONCLUSION:For most practical purposes,CCT measurements with the RTVue,Sirius and Lenstar can be used interchangeably.Although highly correlated,CCTmeasurement differences between USP and these 3optical instruments can be significant depending on the clinical situation.展开更多
文摘AIM:To compare the RTVue spectral optical coherence tomography(SD-OCT),Sirius Scheimpflug-Placido topographer,Lenstar optical low coherence reflectometry(OLCR)and ultrasound pachymetry(USP)devices in terms of their agreement and repeatability of measuring central corneal thickness(CCT).METHODS:In this prospective study,50 eyes of 50patients were included.Three repeated measures were obtained using SD-OCT,Scheimpflug-Placido topographer and USP and five measurements were determined with the OLCR.Bland-Altman plots were used to assess agreement among the instruments,and 95%limits of agreement(LoA)for each comparison were calculated.Intra-examiner repeatability was assessed using intraclass correlation coefficients(ICCs).RESULTS:The mean CCT by SD-OCT,ScheimpflugPlacido topographer,OLCR,and USP were 525.90±34.08μm,525.92±34.10μm,530.30±35.62μm,and543.50±37.11μm respectively.All 4 modalities of CCT measurements correlated closely with each other,with Pearson correlation coefficients ranging from 0.977 to0.995.The mean differences(and upper/lower LoA)for CCT measurements were-0.05±6.77μm(13.3/-13.3)between SD-OCT and Scheimpflug-Placido topographer,4.38±3.79μm(11.8/-3.1)between OLCR and SD-OCT,4.38±6.03μm(16.2/-7.5)between OLCR and ScheimpflugPlacido topographer,13.20±6.46μm(25.9/0.5)between USP and OLCR,17.59±6.76μm(30.8/4.3)between USP and SD-OCT,and 17.58±8.13μm(33.5/1.6)between USP and Scheimpflug-Placido topographer.Intra-examiner repeatability was excellent for all devices with ICCs】0.98.CONCLUSION:For most practical purposes,CCT measurements with the RTVue,Sirius and Lenstar can be used interchangeably.Although highly correlated,CCTmeasurement differences between USP and these 3optical instruments can be significant depending on the clinical situation.