We describe the case of a 77-year-old female who underwent dual-chamber permanent pacemaker implantation using two active screw-in leads for complete atrioventricular block. The two active screw-in leads perforated th...We describe the case of a 77-year-old female who underwent dual-chamber permanent pacemaker implantation using two active screw-in leads for complete atrioventricular block. The two active screw-in leads perforated the atrial and ventricular walls causing cardiac tamponade and a left hemothorax associated with ventricular perforation. An emergency sternotomy was performed to facilitate drainage and hemostasis. Although hemothorax due to pacing lead perforation is very rare, such a possibility should always be considered.展开更多
文摘We describe the case of a 77-year-old female who underwent dual-chamber permanent pacemaker implantation using two active screw-in leads for complete atrioventricular block. The two active screw-in leads perforated the atrial and ventricular walls causing cardiac tamponade and a left hemothorax associated with ventricular perforation. An emergency sternotomy was performed to facilitate drainage and hemostasis. Although hemothorax due to pacing lead perforation is very rare, such a possibility should always be considered.