The dosimetric inter-comparison studies carry pertinent significance to ensure uniform radiation dose delivery for clinical trials. This paper investigates the comparative performance analysis of an X-ray high power l...The dosimetric inter-comparison studies carry pertinent significance to ensure uniform radiation dose delivery for clinical trials. This paper investigates the comparative performance analysis of an X-ray high power linear accelerator per- formed by the International Dosimetry Survey Mission and Institute of Nuclear Medicine & Oncology (INMOL), Lahore. The measurements were made using cylindrical ionization chambers based on the International Atomic Energy Agency (IAEA) TRS-398 protocol for absorbed dose-to-water dosimetric standards, and the percentage deviation was found to be between 0.5 % - 1%. The dosimetric analysis concerning linear X-ray accelerator output performed by INMOL, Lahore was found to be in good agreement with the results of IAEA Dosimtery Survey Mission.展开更多
文摘The dosimetric inter-comparison studies carry pertinent significance to ensure uniform radiation dose delivery for clinical trials. This paper investigates the comparative performance analysis of an X-ray high power linear accelerator per- formed by the International Dosimetry Survey Mission and Institute of Nuclear Medicine & Oncology (INMOL), Lahore. The measurements were made using cylindrical ionization chambers based on the International Atomic Energy Agency (IAEA) TRS-398 protocol for absorbed dose-to-water dosimetric standards, and the percentage deviation was found to be between 0.5 % - 1%. The dosimetric analysis concerning linear X-ray accelerator output performed by INMOL, Lahore was found to be in good agreement with the results of IAEA Dosimtery Survey Mission.