The present study utilized a porcine model for qualitative and quantitative assessment of the diagnostic quality of non-contrast abdominal computed tomography (CT) images generated by Adaptive Statistical Iterative Re...The present study utilized a porcine model for qualitative and quantitative assessment of the diagnostic quality of non-contrast abdominal computed tomography (CT) images generated by Adaptive Statistical Iterative Reconstruction (ASIR, GE Healthcare, Waukesha, Wisconsin, USA), Model-Based Iterative Reconstruction (GE company name VEO), and conventional Filtered back projection (FBP) technique. Methods: Multiple CT whole-body scans of a freshly euthanized pig carcass were performed on a 64-slice GE CT scanner at varying noise indices (5, 10, 15, 20, 30, 37, 40, 45), and with three different algorithms (VEO, FBP, and ASIR at 30%, 50%, and 70% levels of ASIR-FBP blending). Abdominal CT images were reviewed and scored in a blinded and randomized manner by two board-certified abdominal radiologists. The task was to evaluate the clarity of the images according to a rubric involving edge sharpness, presence of artifact, anatomical clarity (assessed at four regions), and perceived diagnostic acceptability. This amounted to seven criteria, each of which was graded on a scale of 1 to 5. A weighted formula was used to calculate a composite score for each scan. Results: VEO outperforms ASIR and FBP by an average of 0.5 points per the scoring system used (p < 0.05). Above a threshold noise index of 30, diagnostic acceptability is lost by all algorithms, and there is no diagnostic advantage to increasing the dose beyond a noise index of 10. Between a noise index of 25 - 30, VEO retains diagnostic acceptability, as opposed to ASIR and FBP which lose acceptability above noise index of 25. Conclusion: Model-based iterative reconstruction provides superior image quality and anatomical clarity at reduced radiation dosages, supporting the routine use of this technology, particularly in pediatric abdominal CT scans.展开更多
文摘The present study utilized a porcine model for qualitative and quantitative assessment of the diagnostic quality of non-contrast abdominal computed tomography (CT) images generated by Adaptive Statistical Iterative Reconstruction (ASIR, GE Healthcare, Waukesha, Wisconsin, USA), Model-Based Iterative Reconstruction (GE company name VEO), and conventional Filtered back projection (FBP) technique. Methods: Multiple CT whole-body scans of a freshly euthanized pig carcass were performed on a 64-slice GE CT scanner at varying noise indices (5, 10, 15, 20, 30, 37, 40, 45), and with three different algorithms (VEO, FBP, and ASIR at 30%, 50%, and 70% levels of ASIR-FBP blending). Abdominal CT images were reviewed and scored in a blinded and randomized manner by two board-certified abdominal radiologists. The task was to evaluate the clarity of the images according to a rubric involving edge sharpness, presence of artifact, anatomical clarity (assessed at four regions), and perceived diagnostic acceptability. This amounted to seven criteria, each of which was graded on a scale of 1 to 5. A weighted formula was used to calculate a composite score for each scan. Results: VEO outperforms ASIR and FBP by an average of 0.5 points per the scoring system used (p < 0.05). Above a threshold noise index of 30, diagnostic acceptability is lost by all algorithms, and there is no diagnostic advantage to increasing the dose beyond a noise index of 10. Between a noise index of 25 - 30, VEO retains diagnostic acceptability, as opposed to ASIR and FBP which lose acceptability above noise index of 25. Conclusion: Model-based iterative reconstruction provides superior image quality and anatomical clarity at reduced radiation dosages, supporting the routine use of this technology, particularly in pediatric abdominal CT scans.