期刊文献+
共找到1篇文章
< 1 >
每页显示 20 50 100
Alignment of beam position monitors in cryomodule of CADS injector Ⅱ 被引量:2
1
作者 Jian-Dong Yuan Yuan He +7 位作者 Bin Zhang Jun-Hui Zhang Jun-Xia Wu Yong Zhang Feng-Feng Wang jun-jie yao Guo-Zhen Sun Yu-Qin Wan 《Nuclear Science and Techniques》 SCIE CAS CSCD 2017年第6期95-100,共6页
Significant temperature difference(300-77 K or even 4 K) can cause large deformations and displacements of the beam position monitors(BPMs),which affect BPMs measurement resolution or even cause their malfunction in c... Significant temperature difference(300-77 K or even 4 K) can cause large deformations and displacements of the beam position monitors(BPMs),which affect BPMs measurement resolution or even cause their malfunction in cryogenic situations.In this paper,to check the offset from the mechanical to electrical center in low temperature(77 K),Fourier's law and finite element method are used to simulate cryo-deformation.Laser tracker and micro-alignment telescope are employed in combined BPM calibration,installation and monitoring.The calibration error is<0.02 mm,and the installation and monitoring precision are 0.06 mm and 0.01 mm,respectively.The monitored cryo-deformation agrees well with the simulation results.These indicate that the combined alignment can improve performance of the BPM system.All these guaranteed the success of running the 9.55 MeV@2.14 mA cw protons. 展开更多
关键词 BEAM POSITION MONITORS ALIGNMENT Fourier’s LAW Cryomodule CADS ANSYS