New technique to control in situ oxygen fugacity in water-free high-pressure system──Using Cu-O as an example
1
作者 li heping 1,{1., guo jie 1, zhang yueming 1,xu zuming 1 and xu ji’an 2 1. institute of geochemistry, chinese academy of sciences, guiyang 550002, china 2. institute of Earth sciences, Taiwan 11529, china 《Chinese Science Bulletin》 SCIE CAS 1998年第16期1353-1358,共6页
Through inserting a solid oxygen-specific electrolyte (YSZ) inside and exerting an electrical power of direct current outside between the sample and oxygen reservoir (an solid oxygen buffer), oxygen was driven by the ... Through inserting a solid oxygen-specific electrolyte (YSZ) inside and exerting an electrical power of direct current outside between the sample and oxygen reservoir (an solid oxygen buffer), oxygen was driven by the power from the cathodic electrode area into the positive electrode area at high temperature and high pressure. In this way, the oxygen fugacity in the sample was in situ controlled independently of temperature and pressure. By using Cu-O system as an example, this technique was proved to be successful. 展开更多
关键词 in SITU control of oxygen FUGACITY water-free SYSTEM HIGH temperature and HIGH pressure.