Electrochemical behavior of dental implant system before and after clinical use
1
作者 Chae-Heon CHUNG Hee-Jung KIM +3 位作者 Yong-Tae JEONG mee-kyoung son Yong-Hoon JEONG Han-Cheol CHOE 《中国有色金属学会会刊:英文版》 EI CSCD 2009年第4期846-851,共6页
Electrochemical behavior of dental implant system before and after clinical use(in vivo and in vitro) was researched by using abutment and titanium fixture.To simulate an oral environment,the samples of clinically use... Electrochemical behavior of dental implant system before and after clinical use(in vivo and in vitro) was researched by using abutment and titanium fixture.To simulate an oral environment,the samples of clinically used and non-used implant systems as a working electrode were exposed to artificial saliva at(36.5±1) ℃.Electrochemical tests were carried out using a potentiostat.After electrochemical test,the corrosion morphology of each sample was investigated by FE-SEM and EDS.The corrosion potential and pitting potential of clinically used implant system are lower than those of non-used implant system,and clinically used implant system exhibits a lower range of passivation,indicating a less degree of inherent resistance against chloride ion.The polarization resistance decreases in the case of clinically used implant system,whereas,Rp for clinically non-used implant system increases compared with clinically used implant system. 展开更多
关键词 电化学行为 种植制度 临床应用 使用制度 临床使用 电化学测试 夹具系统 腐蚀形态