Systematical Experiment for Optimal Design of Vibrating Flow Pump with Jelly-fish Valve 被引量:1
1
作者 Satoyuki Kawano saki miyagawa Atsushi Shirai 《Journal of Bionic Engineering》 SCIE EI CSCD 2016年第1期166-179,共14页
The fundamental characteristics and the flow mechanism of a Vibrating Flow Pump (VFP) with a jelly-fish valve, which can be applied to a novel artificial heart, were studied theoretically and experimentally. By usin... The fundamental characteristics and the flow mechanism of a Vibrating Flow Pump (VFP) with a jelly-fish valve, which can be applied to a novel artificial heart, were studied theoretically and experimentally. By using water as the working fluid, the measurement methodology for the typical unsteady flow for VFP was developed here. The effects of the frequency, amplitude and inner diameter of the vibrating pipe, and thickness of the silicone rubber sheet of the jelly-fish valve on the basic per- formance of VFP were systematically investigated. A high-speed observation technique and simple theoretical model analysis were also introdueed for further detailed discussion. Quantitative contributions of the individual parameters to the pumping performance were shown through the experiment, which would give us essential knowledge for establishing design criteria of VFE The theoretical model, which agreed with the experiment and the high-speed observation, elucidated the pumping mechanism with respect to the role of inertia of the inner fluid. 展开更多
关键词 vibrating flow pump artificial heart left ventricular assistance jelly-fish valve pump performance H-Q curve