期刊文献+
共找到1篇文章
< 1 >
每页显示 20 50 100
Removal of Alcyon II, CGR, MeV <sup>60</sup>Co Teletherapy Head and Evaluation of Exposure Dose
1
作者 Yasser Tawflk Selim Yasser Fouad Lasheen +1 位作者 Mahmoud Ali Hassan tarik mohamed el zakla 《Journal of Environmental Protection》 2013年第12期1435-1440,共6页
Teletherapy units containing 60Co source are used in Egypt and worldwide for treatment of cancer diseases. One of the widely used in Egypt is Alcyon CGR MeV. Different radiotherapy facilities need to unload this 60Co ... Teletherapy units containing 60Co source are used in Egypt and worldwide for treatment of cancer diseases. One of the widely used in Egypt is Alcyon CGR MeV. Different radiotherapy facilities need to unload this 60Co unit after being disused. The head containing the source was removed and put into approved package for transportation, after the source kept in safe position. This study describes the planning and removing process of the disused sources and, transports it into its head. The personal doses measured during the process. This work pre-estimates the doses received to workers during the removing process. The workers received doses of approximately 55 μSv during the whole process. This study shows that the original lead head can be used as the source container of this 60Co unit but not as a long term storage option. The 60Co machine was smoothly dismantled and transported by trained workers and driver, using planned and controlled radiation protection, followed by ALARA (as low as reasonably achievable) rule. 展开更多
关键词 Alcyon 60Co Teletherapy HEAD Sealed SOURCE
下载PDF
上一页 1 下一页 到第
使用帮助 返回顶部