作者 Maximiliano S.Sangoi vítor todeschini Martin Steppe 《Journal of Pharmaceutical Analysis》 CAS 2015年第2期137-141,共5页
A dissolution test for fesoterodine low dose extended-release tablets using liquid chromatographic(LC) method equipped with a C18 monolithic column was developed and validated. LC system was operated isocratically a... A dissolution test for fesoterodine low dose extended-release tablets using liquid chromatographic(LC) method equipped with a C18 monolithic column was developed and validated. LC system was operated isocratically at controlled temperature(40 1C) using a mobile phase of acetonitrile:methanol:0.03 M ammonium acetate(p H 3.8)(30:15:55, v/v/v), run at a flow rate of 1.5 m L/min and detected at 208 nm. The best dissolution conditions for this formulation were achieved using a USP apparatus 2(paddle) at 100 rpm and 900 m L of phosphate buffer at p H 6.8 as the dissolution medium.Validation parameters such as the specificity, linearity, accuracy, precision, and robustness were evaluated according to international guidelines, giving results within the acceptable range. The kinetic parameters of drug release were also investigated using model-dependent methods and the dissolution profiles were best described by the Higuchi model. The validated dissolution test can be applied for quality control of this formulation. 展开更多
关键词 FESOTERODINE Dissolution test Low dose extendedrelease tablets Monolithic LC Release kinetics