期刊文献+
共找到2篇文章
< 1 >
每页显示 20 50 100
Ocular Toxoplasmosis: A Case Report
1
作者 Sara Ortiz Raul Martin victoria de juan 《International Journal of Clinical Medicine》 2013年第1期1-4,共4页
Report a case of a male patient with a macular scar compatible with ocular toxoplasmosis (OT) in right eye (RE) and review the relevant literature on this disease. A patient, who attended for a routine contact lens fo... Report a case of a male patient with a macular scar compatible with ocular toxoplasmosis (OT) in right eye (RE) and review the relevant literature on this disease. A patient, who attended for a routine contact lens follow up, presented with amblyopic exotropia without any ocular disease. Best-corrected visual acuity of the affected eye was 20/40 with constant and mono-fixation exotropia. Ophthalmoscopic assessment revealed a macular scar compatible with OT. OT is the leading cause of infection in the posterior segment. Inactive cases could be asymptomatic and diagnosis requires a complete eye examination, providing a correct diagnosis and patient management. 展开更多
关键词 TOXOPLASMOSIS RETINOCHOROIDITIS OPHTHALMOSCOPY POSTERIOR UVEITIS
下载PDF
Repeatability of 3D OCT Pachymetry in Eyes with and without Contact Lens Induced Corneal Swelling
2
作者 victoria de juan Ana B. Cisneros Raul Martin 《Advances in Computed Tomography》 2013年第2期63-67,共5页
Introduction: Corneal thickness measurement is an indirect assessment of the physiological corneal function in human eye. This value, especially in the central area, is an important data for clinical diagnosis, treatm... Introduction: Corneal thickness measurement is an indirect assessment of the physiological corneal function in human eye. This value, especially in the central area, is an important data for clinical diagnosis, treatment and monitoring of anterior segment pathologies. There are several technologies for non-invasively measuring corneal thickness like optical coherence tomography (OCT). However, there is little information available describing the effect of contact lens induced corneal swelling in OCT measures. The purpose of this study was to evaluate the repeatability of OCT corneal pachymetry in eyes with and without contact lens induced corneal swelling. Material and Methods: This study included five visits over one week in six healthy subjects: baseline and after sleeping with four different CL of +0.50 D, +2.00 D, +5.00 D and +8.00 D (Acuvue 2, Johnson & Johnson Vision Care) on four different days. Corneal pachymetry OCT 3D (Topcon) were measured three consecutive times during each visit. Besides, this measurement was repeated in 12 meridians (Scans #1 to #12), between 0°and 165°with 15° intervals. Results: Corneal pachymetry was significant different between before and after contact lens wear (p < 0.001 ANOVA). High repeatability was found without corneal swelling [Coefficient of variation (CV) = 0.68% p = 0.93] and with corneal swelling [CL -0.50 D (CV = 0.78% p = 0.95;590 ± 46 μm CI 95% 548 to 596), with CL + 2.00 D (CV = 0.72% p = 0.97;601 ± 46 μm CI 95% 595 to 607), with CL + 5.00 D (CV = 0.78% p = 0.66;608 ± 50 μm CI 95% 601 to 615) and with CL + 8.00 D (CV = 0.77% p = 0.97;607 ± 44 μm CI 95% 601 to 613)]. There were no differences (p = 1.00) in central corneal thickness along the 12 scans corneal [Scan #0 (593 ± 50 μm CI 95% 582 to 603), #1 (592 ± 50 μm CI 95% 581 to 603), #2 (591 ± 50 μm IC 95% 580 to 602), #3 (590 ± 49 μm CI 95% 579 to 600), #4 (590 ± 50 μm CI 95% 579 to 600), #5 (591 ± 50 μm CI 95% 581 to 602), #6 (590 ± 51 μm CI 95% 580 to 601), #7 (589 ± 50 μm CI 95% 578 to 600), #8 (590 ± 50 μm CI 95% 579 to 600), #9 (589 ± 50 μm CI 95% 579 to 600), #10 (592 ± 50 μm CI 95% 581 to 602) and #11 (591 ± 50 μm CI 95% 580 to 601). Conclusions: 3D OCT showed a high repeatability (CV < 1%) to measure central thickness in corneas with and without contact lens induced corneal swelling. This technology presents advantages over other clinical devices used to measure corneal thickness and could be the “gold standard” for future clinical studies and clinical practice for eye care practitioners. 展开更多
关键词 Optical COHERENCE Tomography CORNEAL SWELLING Contact Lenses CORNEAL PACHYMETRY
下载PDF
上一页 1 下一页 到第
使用帮助 返回顶部