Scan system for arbitrary-shaped samples at the synchrotron radiation facility
1
作者 xu-ying lan Dong-Xu Liang Cheng-Wen Mao 《Nuclear Science and Techniques》 SCIE CAS CSCD 2017年第5期18-27,共10页
X-ray fluorescence(XRF) scan methodology is important for elemental mapping of samples at a synchrotron radiation facility. To save the experiment time and improve the experiment efficiency, one should develop an effi... X-ray fluorescence(XRF) scan methodology is important for elemental mapping of samples at a synchrotron radiation facility. To save the experiment time and improve the experiment efficiency, one should develop an efficient XRF scan method. In this paper, a new scan mode is presented. It can map arbitrary-shaped areas withou stopping the motors. The control and data acquisition system integrates motor controlling, detector triggering, and data acquisition and storage. The system realizes the arbitrary-shaped 2D-mapping and fluorescence data acquisition synchronously. SR-XRF mapping has been performed with a standard gold mask to verify the validity of this method a beamline BL15U1 of the Shanghai Synchrotron Radiation Facility. The results show that this method reduces the tota scan time and improves the experiment efficiency. 展开更多
关键词 SYNCHROTRON radiation X-ray fluorescence mapping EPICS XPS controller