Seismometer self-noise levels were determined using Sleeman’s three-sensor method in combination with the Welch method for different parameter combinations. The self-noise levels decreased with the increasing segment...Seismometer self-noise levels were determined using Sleeman’s three-sensor method in combination with the Welch method for different parameter combinations. The self-noise levels decreased with the increasing segment window length, which is equivalent to the subwindow length, and with the increasing segment overlap rate for different frequency points at a fixed band. After the statistical examination of 9800 different parameter combinations, a zone of reasonable self-noise calculation parameter combinations was identified. Reasons for the unsuitability of certain parameter combinations were explored with respect to their distortion of the seismometer’s self-noise levels.展开更多
According to different testing purposes, methods and available environmental conditions, the seismograph testing can be divided into laboratory and on-site testing, respectively. The testing of the seismograph's k...According to different testing purposes, methods and available environmental conditions, the seismograph testing can be divided into laboratory and on-site testing, respectively. The testing of the seismograph's key parameters and other concerning technical specifications are well described in guide documents(China Earthquake Administration, 2017). This includes seismometer sensitivity, linearity and clip levels based on the shake table test, as well as the seismometer natural period, damping constant based on electrical calibration(Wang Guangfu,1986; Ple?inger A.,1993) and instrumental self-noise collocation estimation(Holcomb L.G., 1989; Sleeman R. et al., 2006). However, with the development of seismic observation technology, many new requirements for the performance evaluation of seismographs have been put forward, and new testing items and methods have emerged.展开更多
基金supported by Study on Seismograph Calibration and Preparation of Draft International Standard (201408005)the Earthquake Science and technology Program of Hebei Province (201303)Special Program for Earthquake monitoring (20130203)
文摘Seismometer self-noise levels were determined using Sleeman’s three-sensor method in combination with the Welch method for different parameter combinations. The self-noise levels decreased with the increasing segment window length, which is equivalent to the subwindow length, and with the increasing segment overlap rate for different frequency points at a fixed band. After the statistical examination of 9800 different parameter combinations, a zone of reasonable self-noise calculation parameter combinations was identified. Reasons for the unsuitability of certain parameter combinations were explored with respect to their distortion of the seismometer’s self-noise levels.
基金sponsored by the Department of Earthquake Monitoring and Prediction,China Earthquake Administration
文摘According to different testing purposes, methods and available environmental conditions, the seismograph testing can be divided into laboratory and on-site testing, respectively. The testing of the seismograph's key parameters and other concerning technical specifications are well described in guide documents(China Earthquake Administration, 2017). This includes seismometer sensitivity, linearity and clip levels based on the shake table test, as well as the seismometer natural period, damping constant based on electrical calibration(Wang Guangfu,1986; Ple?inger A.,1993) and instrumental self-noise collocation estimation(Holcomb L.G., 1989; Sleeman R. et al., 2006). However, with the development of seismic observation technology, many new requirements for the performance evaluation of seismographs have been put forward, and new testing items and methods have emerged.