A compact, low cost, multipoint Thomson scattering diagnostic system for HT-7 superconducting tokamak has been in operation since 1999. Its capability of measuring electron temperatures is in the range of 200 eV to 2 ...A compact, low cost, multipoint Thomson scattering diagnostic system for HT-7 superconducting tokamak has been in operation since 1999. Its capability of measuring electron temperatures is in the range of 200 eV to 2 keV at a density of a few times IO12 cm-3, with a spatial resolution of 2.4 cm for 5 spatial points and a temporal resolution of 1 ms-1 s for 8 time points. The main components of the diagnostic system include a 20-25 J Nd: glass laser with 35 ns pulse width (8 pulses per burst), a KDP frequency-doubling unit, spherical mirrors of multipass input optical system, a wide-angle collection objective, a bandpass glass filter for reducing the stray light to zero a f/2.5 polychromator, a fiberglass collimator, a photomultiplier's box with electronic preamplifier, high gain and high signal/noise ratio, CAMAC data acquisition and so on. The multipass optical system has been successful at increasing the quantity of scattered photons by passing the probing laser beam 10 times through the plasma under investigation. The HT7 Thomson scattering diagnostic has provided successfully the information on two-dimensional electron temperature in the plasma of HT-7 tokamak with LHCD and IBW.展开更多
文摘A compact, low cost, multipoint Thomson scattering diagnostic system for HT-7 superconducting tokamak has been in operation since 1999. Its capability of measuring electron temperatures is in the range of 200 eV to 2 keV at a density of a few times IO12 cm-3, with a spatial resolution of 2.4 cm for 5 spatial points and a temporal resolution of 1 ms-1 s for 8 time points. The main components of the diagnostic system include a 20-25 J Nd: glass laser with 35 ns pulse width (8 pulses per burst), a KDP frequency-doubling unit, spherical mirrors of multipass input optical system, a wide-angle collection objective, a bandpass glass filter for reducing the stray light to zero a f/2.5 polychromator, a fiberglass collimator, a photomultiplier's box with electronic preamplifier, high gain and high signal/noise ratio, CAMAC data acquisition and so on. The multipass optical system has been successful at increasing the quantity of scattered photons by passing the probing laser beam 10 times through the plasma under investigation. The HT7 Thomson scattering diagnostic has provided successfully the information on two-dimensional electron temperature in the plasma of HT-7 tokamak with LHCD and IBW.