The differential potentiometric stripping analysis (DPSA) with Nafion coated carbon fibre electrode has been studied, and a method to determine lead and cadmium in urine directly has been attempted. The effect of vari...The differential potentiometric stripping analysis (DPSA) with Nafion coated carbon fibre electrode has been studied, and a method to determine lead and cadmium in urine directly has been attempted. The effect of various experimental parameters on the DPSA response is discussed. The experimental conditions include 0.2M sodium perchlorate, deposition potential of-1.0 or -1.1 V and using 20 ppm mercuric ion as oxidizing agent. The response of the signal is in linear relation with the concentrations of lead and cadmium respectively up to 0.5 ppm. The electrode coated with Nafion film alleviates the interference from organics in urine samples.展开更多
基金This project was supported by the National Natural Science Foundation of China
文摘The differential potentiometric stripping analysis (DPSA) with Nafion coated carbon fibre electrode has been studied, and a method to determine lead and cadmium in urine directly has been attempted. The effect of various experimental parameters on the DPSA response is discussed. The experimental conditions include 0.2M sodium perchlorate, deposition potential of-1.0 or -1.1 V and using 20 ppm mercuric ion as oxidizing agent. The response of the signal is in linear relation with the concentrations of lead and cadmium respectively up to 0.5 ppm. The electrode coated with Nafion film alleviates the interference from organics in urine samples.