期刊文献+
共找到1篇文章
< 1 >
每页显示 20 50 100
EFFICIENT FITTING RANGE FOR GLUCOSE MEASUREMENT WITH OPTICAL COHERENCE TOMOGRAPHY 被引量:1
1
作者 SHU FENG KEHONG YUAN DATIAN YE 《Journal of Innovative Optical Health Sciences》 SCIE EI CAS 2009年第1期107-115,共9页
Studies of non-invasive glucose measurement with optical coherence tomography(OCT)in tissue-simulating phantoms and biological tissues show that glucose has an effect on the OCT signal slope.Choosing an efficient fitt... Studies of non-invasive glucose measurement with optical coherence tomography(OCT)in tissue-simulating phantoms and biological tissues show that glucose has an effect on the OCT signal slope.Choosing an efficient fitting range to calculate the OCT signal slope is important because it helps to improve the precision of glucose measurement.In this paper,we study the problem in two ways:(1)scattering-induced change of OCT signal slope versus depth in intralipid suspensions with different concentrations based on Monte Carlo(MC)simulations and experiments and(2)efficient fitting range for glucose measurement in 3%and 10%intralipid.The results show that the OCT signal slope expresses a contrary change with scattering coefficient below a certain depth in high intralipid concentrations,so that there is an effective fitting depth.With an efficient fitting range from 100μm to the effective fitting depth,the precision of glucose measurement can be 4.4mM for 10%intralipid and 2.2mM for 3%intralipid. 展开更多
关键词 Optical coherence TOMOGRAPHY noninvasive glucose measurement multiple scattered photons
下载PDF